文献速递技术与安全腹部US

腹部体壁对诊断超声原位峰值稀疏压影响的定量研究

Quantifying the Effect of Abdominal Body Wall on In Situ Peak Rarefaction Pressure During Diagnostic Ultrasound Imaging.
PubMed 文献 · Ultrasound Med Biol · 2021-03-13

本研究利用三维非线性超声模拟与实验测量,估计经腹成像中可出现的原位压力范围,并识别线性降额法估算原位峰值稀疏压(PRP)的误差来源。结果显示,大发射孔径(F/1.5)时MI持续高估原位PRP 20%-48%,中孔径(F/3)时MI估算误差在8%以内,小孔径(F/5)时MI持续低估原位PRP 32%-50%。体壁样本与聚焦配置间的大变异性表明简化线性降额方案的局限性,提示患者特异性原位PRP估算可能允许提高发射压力以改善图像质量并确保安全。

要点

  • MI指南的线性降额法在大孔径(F/1.5)时高估原位PRP 20%-48%,主要因相位畸变。
  • 小孔径(F/5)时MI低估原位PRP 32%-50%,峰值位置常位于体壁内、焦深前1-2 cm。
  • 体壁样本与聚焦配置间变异性大,提示患者特异性原位PRP估算可能改善安全性与图像质量。

诊断标准与关键数值

  • 大发射孔径(F/1.5)时,MI持续高估原位PRP 20%-48%,主要归因于相位畸变。
  • 中发射孔径(F/3)时,MI估算原位PRP的误差在8%以内。
  • 小发射孔径(F/5)时,MI持续低估原位PRP 32%-50%,峰值位置出现在焦深前1-2 cm,常位于体壁内。

临床应用与阅片要点

  • 经腹超声成像中,使用大孔径(F/1.5)时MI可能高估原位PRP,需注意实际压力可能低于MI指示值。
  • 使用小孔径(F/5)时MI可能低估原位PRP,且峰值位置常位于体壁内、焦深前1-2 cm,需关注体壁处的压力峰值。
  • 体壁样本与聚焦配置间变异性大,提示简化线性降额方案在不同患者和配置下可能不准确。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

In this study, 3-D non-linear ultrasound simulations and experimental measurements were used to estimate the range of in situ pressures that can occur during transcutaneous abdominal imaging and to identify the sources of error when estimating in situ peak rarefaction pressures (PRPs) using linear derating, as specified by the mechanical index (MI) guideline. Using simulations, it was found that, for a large transmit aperture (F/1.5), MI consistently over-estimated in situ PRP by 20%-48% primarily owing to phase aberration. For a medium transmit aperture (F/3), the MI accurately estimated the in situ PRP to within 8%. For a small transmit aperture (F/5), MI consistently underestimated the in situ PRP by 32%-50%, with peak locations occurring 1-2 cm before the focal depth, often within the body wall itself. The large variability across body wall samples and focal configurations demonstrates the limitations of the simplified linear derating scheme. The results suggest that patient-specific in situ PRP estimation would allow for increases in transmit pressures, particularly for tightly focused beams, to improve diagnostic image quality while ensuring patient safety.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Zhang B, Pinton GF, Deng Y, et al. Quantifying the Effect of Abdominal Body Wall on In Situ Peak Rarefaction Pressure During Diagnostic Ultrasound Imaging. Ultrasound Med Biol. 2021. PMID: 33722439. doi:10.1016/j.ultrasmedbio.2021.01.028

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。