文献速递心血管外周血管技术与安全CT

光子计数多能CT联合光谱优化对比剂用于斑块去除与狭窄评估

Photon-Counting Multienergy Computed Tomography With Spectrally Optimized Contrast Media for Plaque Removal and Stenosis Assessment.
PubMed 文献 · Invest Radiol · 2021-01-01

在冠状动脉狭窄体模中,将5种研究性对比剂(碘、铋、铪、钬、钨)与光子计数探测器CT的光谱优化能量阈值结合,通过线性双物质分解重建对比剂图以减除钙化斑块。钨和铪的对比剂图噪声低于碘,钨改善管腔显示并提高可高置信度评估的狭窄程度。

要点

  • 钨对比剂图平均噪声49 HU,低于碘的78 HU,管腔显示优于碘。
  • 钨、铋、铪可高置信度评估的狭窄程度中位数分别为70%、70%、75%,碘为50%。
  • 钬对比剂图噪声165 HU,高于碘,管腔显示较碘下降。

关键影像表现/征象

  • 对比剂图平均噪声:钨49 HU、铪62 HU低于碘78 HU;铋81 HU、钬165 HU高于碘。
  • 各对比剂图噪声纹理相似。
  • 短轴对比剂图管腔显示读者间一致性k=0.567–0.814。
  • 与碘相比,钨改善管腔显示(两位读者均P<0.001),铪类似改善(P=0.008、P=0.29),铋类似(P=0.38、P=0.69),钬下降(均P<0.001)。
  • 整体主观图像质量:钬与碘相似,钨、铪、铋优于碘。

诊断标准与关键数值

  • 以碘为参考标准,两位盲读者独立评分短轴管腔显示和长轴整体主观图像质量。
  • 四位读者确定长轴对比剂图上可高诊断置信度评估的最高狭窄程度:钨均值70%(IQR 70%–70%)、铋70%(60%–70%)、铪75%(70%–80%),碘50%(50%–60%)、钬50%(50%–60%)。

临床应用与阅片要点

  • 阅读PCCT对比剂图时,钨和铪的图像噪声较低,可能更利于小血管管腔显示。
  • 钬对比剂图噪声较高且管腔显示较碘下降,评估时需注意。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The aim of this study was to systematically evaluate the potential to combine investigational contrast media with spectrally optimized energy-thresholding of photon-counting detector computed tomography (PCCT) for subtraction of calcified plaques in a coronary artery stenosis phantom.

METHODS: A small vessel phantom containing 3 fillable tubes (diameter, 3 mm each) with calcified plaques was placed into an anthropomorphic chest phantom. The plaques had incremental thicknesses ranging from 0.3 to 2.7 mm, simulating vessel stenoses ranging from 10% to 90% of the lumen diameter. The phantom was filled with 5 different investigational contrast media (iodine, bismuth, hafnium, holmium, and tungsten) at equal mass concentrations (15 mg/mL) and was imaged on a prototype PCCT at 140 kVp using optimized, contrast media-dependent energy thresholds. Contrast maps (CMs) were reconstructed for each contrast medium by applying a linear 2-material decomposition algorithm. Image noise magnitude and noise texture of CM were compared among the contrast media using the noise power spectrum. Two blinded readers independently rated the vessel lumen visualization on short-axis and the overall subjective image quality on long-axis CM relative to iodine as the reference standard. Four readers determined the highest degree of stenosis that could be assessed with high diagnostic confidence on long-axis CM.

RESULTS: Average image noise on CM was lower for tungsten (49 HU) and hafnium (62 HU) and higher for bismuth (81 HU) and holmium (165 HU) compared with iodine (78 HU). Noise texture of CM was similar among the contrast media. Interreader agreement for vessel lumen visualization on short-axis CM ranged from moderate to excellent (k = 0.567-0.814). Compared with iodine, lumen visualization of each reader was improved using tungsten (P < 0.001 for both readers), similar to improved using hafnium (P = 0.008, P = 0.29), similar using bismuth (P = 0.38, P = 0.69), and decreased using holmium (both, P < 0.001). Overall subjective image quality was similar for holmium and superior for tungsten, hafnium, and bismuth as compared with iodine. Higher-degree stenoses were evaluable with high confidence using tungsten (mean, 70%; interquartile range, 70%-70%), bismuth (70%; 60%-70%), and hafnium (75%; 70%-80%) compared with iodine (50%; 50%-60%) and holmium (50%; 50%-60%).

CONCLUSIONS: Spectral optimization in PCCT combined with investigational contrast media can improve calcium subtraction and stenosis assessment in small vessels. Contrast maps of tungsten and, to a lesser extent, hafnium as contrast media yielded superior image noise properties and improved vessel lumen visualization, along with a higher subjective image quality compared with the reference standard iodine.

来源:Invest Radiol,查看原文。版权归原作者所有。