单次屏气与自导航超短回波时间序列肺功能MRI评估
前瞻性探索性研究纳入10例患者和10名健康对照,在3.0T下比较屏气stack-of-spirals与自由呼吸自导航Koosh ball两种3D UTE序列的图像质量和功能分析适用性。两种技术图像质量评分相同,但屏气序列更易受运动伪影影响;屏气时FVSI均值高于自由呼吸,两种技术AVF相关性好。
要点
- 两种3D UTE序列图像质量评分相同,均可用于质子密度加权肺功能分析。
- 屏气stack-of-spirals序列更易出现运动和混叠伪影。
- 屏气时FVSI均值高于自由呼吸,两种技术AVF相关性好。
关键影像表现/征象
- 屏气stack-of-spirals 3D UTE序列(TE 0.05 msec)比自由呼吸自导航Koosh ball 3D UTE序列(TE 0.03 msec)更易受运动和混叠伪影影响。
- 两种技术图像质量四点Likert评分相同。
诊断标准与关键数值
- 图像质量采用四点Likert量表评分,两种技术评分相同。
- 屏气FVSI均值0.17 mL/mL±0.06(最小0.07,最大0.34) vs 自由呼吸0.11 mL/mL±0.03(最小0.06,最大0.17),P=0.016。
- FVSI与FVVol平均差异:屏气-0.008(95%CI 0.007,-0.024),ρ=0.97;自由呼吸-0.004(95%CI 0.007,-0.016),ρ=0.91。
- AVF在两种技术间相关ρ=0.94。
临床应用与阅片要点
- 屏气序列需注意运动和混叠伪影对图像质量的影响。
- 两种序列均可用于半自动分割后计算FVSI、FVVol和AVF等功能指标。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate three-dimensional (3D) ultrashort echo time (UTE) MRI regarding image quality and suitability for functional image analysis using gradient-echo sequences in breath-hold and with self-navigation.
MATERIALS AND METHODS: In this prospective exploratory study, 10 patients (mean age, 21 years; age range, 5-58 years; five men) and 10 healthy control participants (mean age, 25 years; age range, 10-39 years; five men) underwent 3D UTE MRI at 3.0 T. Imaging was performed with a prototypical stack-of-spirals 3D UTE sequence during single breath holds (echo time [TE], 0.05 msec) and with a self-navigated "Koosh ball" 3D UTE sequence at free breathing (TE, 0.03 msec). Image quality was rated on a four-point Likert scale. Edge sharpness was calculated. After semiautomated segmentation, fractional ventilation was calculated from MRI signal intensity (FVSI) and volume change (FVVol). The air volume fraction (AVF) was estimated from relative signal intensity (aortic blood signal intensity was used as a reference). Means were compared between techniques and participants. The Wilcoxon signed rank test and Spearman rank correlation were used for statistical analyses.
RESULTS: Image quality ratings were equal for both techniques. However, stack-of-spirals breath-hold UTE was more susceptible to motion and aliasing artifacts. Mean FVSI was higher during breath hold than at free breathing (mean ± standard deviation in milliliters of gas per milliliters of parenchyma, 0.17 mL/mL ± 0.06 [minimum, 0.07; maximum, 0.34] vs 0.11 mL/mL ± 0.03 [minimum, 0.06; maximum, 0.17], P = .016). Mean FVSI and FVVol were in good agreement (mean difference: at breath hold, -0.008 [95% confidence interval {CI}: 0.007, -0.024]; ρ = 0.97 vs free breathing, -0.004 [95% CI: 0.007, -0.016]; ρ = 0.91). AVF correlated between both techniques (ρ = 0.94).
CONCLUSION: Breath-hold and self-navigated 3D UTE sequences yield proton density-weighted images of the lungs that are similar in quality, and both techniques are suitable for functional image analysis.Supplemental material is available for this article.© RSNA, 2020.
引用
- Heidenreich JF, Veldhoen S, Metz C, et al. Functional MRI of the Lungs Using Single Breath-Hold and Self-Navigated Ultrashort Echo Time Sequences. Radiol Cardiothorac Imaging. 2020. PMID: 33778581. doi:10.1148/ryct.2020190162
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。