光子计数探测器CT提高冠状动脉钙化定量准确性:尸体标本离体研究

Improved coronary calcification quantification using photon-counting-detector CT: an ex vivo study in cadaveric specimens.
PubMed 文献 · Eur Radiol · 2021-03-13

该离体研究将离体冠状动脉标本在光子计数探测器CT(PCD-CT)上分别用PCD子系统和能量积分探测器(EID)子系统扫描,并与micro-CT参考标准比较钙化体积。结果显示钙化体积测量趋势为EID-D50 > PCD-D50 > PCD-D60 > micro-CT,各 pairwise 比较均有统计学差异(p < 0.05)。PCD-CT定量冠状动脉钙化比EID-CT更准确,更锐利的PCD-CT卷积核可进一步提高准确性,且PCD-CT图像噪声低于EID-CT。

要点

  • PCD-CT定量冠状动脉钙化体积较EID-CT更接近micro-CT参考标准。
  • 采用更锐利卷积核(D60)的PCD-CT进一步提高了钙化体积定量准确性。
  • 匹配辐射剂量和重建卷积核条件下,PCD-CT图像噪声低于EID-CT。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the accuracy of coronary calcium quantification of cadaveric specimens imaged from a photon-counting detector (PCD)-CT and an energy-integrating detector (EID)-CT.

METHODS: Excised coronary specimens were scanned on a PCD-CT scanner, using both the PCD and EID subsystems. The scanning and reconstruction parameters for EID-CT and PCD-CT were matched: 120 kV, 9.3-9.4 mGy CTDIvol, and a quantitative kernel (D50). PCD-CT images were also reconstructed using a sharper kernel (D60). Scanning the same specimens using micro-CT served as a reference standard for calcified volumes. Calcifications were segmented with a half-maximum thresholding technique. Segmented calcified volume differences were analyzed using the Friedman test and post hoc pairwise Wilcoxon signed rank test with the Bonferroni correction. Image noise measurements were compared between EID-CT and PCD-CT with a repeated-measures ANOVA test and post hoc pairwise comparison with the Bonferroni correction. A p < 0.05 was considered statistically significant.

RESULTS: The volume measurements in 12/13 calcifications followed a similar trend: EID-D50 > PCD-D50 > PCD-D60 > micro-CT. The median calcified volumes in EID-D50, PCD-D50, PCD-D60, and micro-CT were 22.1 (IQR 10.2-64.8), 21.0 (IQR 9.0-56.5), 18.2 (IQR 8.3-49.3), and 14.6 (IQR 5.1-42.4) mm3, respectively (p < 0.05 for all pairwise comparisons). The average image noise in EID-D50, PCD-D50, and PCD-D60 was 60.4 (± 3.5), 56.0 (± 4.2), and 113.6 (± 8.5) HU, respectively (p < 0.01 for all pairwise comparisons).

CONCLUSION: The PCT-CT system quantified coronary calcifications more accurately than EID-CT, and a sharp PCD-CT kernel further improved the accuracy. The PCD-CT images exhibited lower noise than the EID-CT images.

KEY POINTS: • High spatial resolution offered by PCD-CT reduces partial volume averaging and consequently leads to better morphological depiction of coronary calcifications. • Improved quantitative accuracy for coronary calcification volumes could be achieved using high-resolution PCD-CT compared to conventional EID-CT. • PCD-CT images exhibit lower image noise than conventional EID-CT at matched radiation dose and reconstruction kernel.

来源:Eur Radiol,查看原文。版权归原作者所有。