文献速递胸部技术与安全DR

便携式胸部X线检查在COVID-19疫情期间通过玻璃门拍摄的技术与图像质量评估

Social Distancing with Portable Chest Radiographs During the COVID-19 Pandemic: Assessment of Radiograph Technique and Image Quality Obtained at 6 Feet and Through Glass.
PubMed 文献 · Radiol Cardiothorac Imaging · 2020-11-12

本研究评估了在急诊科通过患者房间玻璃门进行便携式胸部X线检查的改良技术,回顾性分析100张X线片(50张改良技术,50张标准技术),由三位胸部放射科医生评估图像质量,并计算患者和工作人员的辐射暴露,同时调查32名X线技术人员的感知。结果显示改良技术图像保持诊断质量,患者辐射剂量相同(169 µGy),技术人员暴露降低(0.055 µGy vs 0.088 µGy),且技术人员接受度高。

要点

  • 改良技术通过玻璃门拍摄的便携式胸片保持诊断质量,多数图像被三位 reviewers 评为可诊断。
  • 患者入口剂量在两种技术间相同,均为169 µGy(50th percentile患者)。
  • 技术人员暴露从标准技术的0.088 µGy降低至改良技术的0.055 µGy(站立于玻璃侧面6英尺处)。

关键影像表现/征象

  • 三位审阅者评估100张便携式胸片,其中两位审阅者认为所有图像具有诊断质量,另一位审阅者认为两张标准技术图像和一张改良技术图像为非诊断性。

临床应用与阅片要点

  • 改良技术可在急诊科通过患者房间玻璃门进行便携式胸部X线检查,图像质量保持诊断水平。
  • 技术人员站立于玻璃侧面6英尺处时暴露剂量为0.055 µGy,低于标准技术的0.088 µGy。
  • 多数技术人员认为改良技术不更难操作,感觉更安全,并确认节省了个人防护装备(PPE)。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To develop a technique that allows portable chest radiography to be performed through the glass door of a patient's room in the emergency department.

MATERIALS AND METHODS: A retrospective review of 100 radiographs (50 [mean age 59.4 ± 17.3, range 22-87; 30 women] performed with the modified technique in April 2020, randomized with 50 [mean age 59 ± 21.6, range 19-100; 31 men] using the standard technique was completed by three thoracic radiologists to assess image quality. Radiation exposure estimates to patient and staff were calculated. A survey was created and sent to 32 x-ray technologists to assess their perceptions of the modified technique. Unpaired Ttests were used for numerical data. A P value < .05 was considered statistically significant.

RESULTS: The entrance dose for a 50th percentile patient was the same between techniques, measuring 169 µGy. The measured technologist exposure from the modified technique assuming a 50th percentile patient and standing 6 feet to the side of the glass was 0.055 µGy, which was lower than standard technique technologist exposure of 0.088 µGy. Of the 100 portable chest radiographs evaluated by three reviewers, two reviewers rated all images as having diagnostic quality, while the other reviewer believed two of the standard images and one of the modified technique images were non-diagnostic. A total of 81% (26 of 32) of eligible technologists completed the survey. Results showed acceptance of the modified technique with the majority feeling safer and confirming conservation of PPE. Most technologists did not feel the modified technique was more difficult to perform.

CONCLUSIONS: The studies acquired with the new technique remained diagnostic, patient radiation doses remained similar, and technologist dose exposure were decreased with modified positioning. Perceptions of the new modified technique by frontline staff were overwhelmingly positive.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。

引用

  1. Gange CP, Pahade JK, Cortopassi I, et al. Social Distancing with Portable Chest Radiographs During the COVID-19 Pandemic: Assessment of Radiograph Technique and Image Quality Obtained at 6 Feet and Through Glass. Radiol Cardiothorac Imaging. 2020. PMID: 33778645. doi:10.1148/ryct.2020200420

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。