呼吸触发2D kat-ARC cine评估法洛四联症术后双心室容积与功能的可靠性
本研究在30例法洛四联症术后患者中比较呼吸触发2D kat-ARC cine与屏气2D SSFP cine测得的双心室容积、功能及图像质量差异。2D kat-ARC扫描时间显著缩短且左心室各项参数无显著差异,但图像质量显著受损,右心室ESV被高估、SV与EF被低估。结论认为2D kat-ARC cine是评估左心室容积与功能的可靠技术。
要点
- 2D kat-ARC扫描时间35.2±9.1 s,显著短于2D SSFP的80.4±16.7 s。
- 图像质量2D kat-ARC较2D SSFP显著受损(p<0.0001)。
- 右心室参数存在差异:ESV被高估,SV与EF被低估。
关键影像表现/征象
- 2D kat-ARC图像质量较2D SSFP cine显著受损(p<0.0001)。
- 双心室EDV、LVESV、LVSV、LVEF及LVM在两种序列间无显著差异。
- RVESV在2D kat-ARC为73.8±43.2 ml,高于2D SSFP的70.3±44.5 ml(p=0.0002)。
- RVSV在2D kat-ARC为46.2±20.5 ml,低于2D SSFP的49.4±20.4 ml(p=0.0024)。
- RVEF在2D kat-ARC为40.2±12.7%,低于2D SSFP的43.5±14.0%(p=0.0002)。
诊断标准与关键数值
- 左心室参数(EDV、ESV、SV、EF、LVM)在2D kat-ARC与2D SSFP间无显著差异。
- 右心室ESV在2D kat-ARC较2D SSFP高估(73.8±43.2 ml vs 70.3±44.5 ml,p=0.0002)。
- 右心室SV在2D kat-ARC较2D SSFP低估(46.2±20.5 ml vs 49.4±20.4 ml,p=0.0024)。
- 右心室EF在2D kat-ARC较2D SSFP低估(40.2±12.7% vs 43.5±14.0%,p=0.0002)。
临床应用与阅片要点
- 2D kat-ARC可显著缩短扫描时间(35.2±9.1 s vs 80.4±16.7 s),适用于左心室容积与功能评估。
- 采用2D kat-ARC评估右心室时需注意其ESV偏高、SV与EF偏低的倾向。
- 2D kat-ARC图像质量较2D SSFP显著受损,阅片时需考虑其对图像判读的影响。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare left ventricular (LV) and right ventricular (RV) volume, function, and image quality of a respiratory-triggered two-dimensional (2D)-cine k-adaptive-t-autocalibrating reconstruction for Cartesian sampling (2D kat-ARC) with those of the standard reference, namely, breath-hold 2D balanced steady-state free precession (2D SSFP), in patients with repaired tetralogy of Fallot (TOF).
METHODS: 30 patients (14 males, mean age 32.2 ± 13.9 years) underwent cardiac magnetic resonance, and 2D kat-ARC and 2D SSFP images were acquired on short-axis view. Biventricular end-diastolic volume (EDV) and end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF), and LV mass (LVM) were analysed.
RESULTS: The 2D kat-ARC had significantly shorter scan time (35.2 ± 9.1 s vs 80.4 ± 16.7 s; p < 0.0001). Despite an analysis of image quality showed significant impairment using 2D kat-ARC compared to 2D SSFP cine (p < 0.0001), the two sequences demonstrated no significant difference in terms of biventricular EDV, LVESV, LVSV, LVEF, and LVM. However, the RVESV was overestimated for 2D kat-ARC compared with that for 2D SSFP (73.8 ± 43.2 ml vs 70.3 ± 44.5 ml, p = 0.0002) and the RVSV and RVEF were underestimated (RVSV = 46.2±20.5 ml vs 49.4 ± 20.4 ml, p = 0.0024; RVEF = 40.2±12.7% vs. 43.5±14.0%, p = 0.0002).
CONCLUSION: Respiratory-triggered 2D kat-ARC cine is a reliable technique that could be used in the evaluation of LV volumes and function.
ADVANCES IN KNOWLEDGE: 2D cine kat-ARC is a reliable technique for the assessment LV volume and function in patients with repaired TOF.
引用
- Orii M, Sugawara T, Takagi H, et al. Reliability of respiratory-triggered two-dimensional cine k-adaptive-t-autocalibrating reconstruction for Cartesian sampling for the assessment of biventricular volume and function in patients with repaired tetralogy of Fallot. Br J Radiol. 2021. PMID: 33733811. doi:10.1259/bjr.20201249
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。