新型平板锥形束CT与多层CT评估急性缺血性卒中的前瞻性对比研究
前瞻性纳入65例大血管闭塞急性缺血性卒中患者,在血管造影室行两种无增强锥形束CT(标准20.8秒18例、改进10.4秒47例),与多层CT对比图像质量。改进10.4秒方案对比度噪声比更高、噪声更低,74.1%扫描实质成像可接受,83%图像质量足以支持直接至血管造影室筛选。
要点
- 改进10.4秒锥形束CT的对比度噪声比显著高于标准20.8秒方案(2.08±0.64 vs 1.15±0.27)。
- 10.4秒方案74.1%扫描实质成像评为可接受,标准方案仅32.4%,多层CT为100%。
- 83%的10.4秒锥形束CT图像质量足以支持直接至血管造影室筛选,标准方案仅11%。
关键影像表现/征象
- 10.4秒方案的平均对比度噪声比为2.08±0.64,标准20.8秒方案为1.15±0.27。
- 10.4秒方案的平均图像噪声为6.30±1.34,标准20.8秒方案为7.82±2.03。
- 图像质量改善最大的方面包括灰白质分化(改善59%)、图像噪声降低(改善63%)和伪影减少(改善50%)。
诊断标准与关键数值
- 定性分析采用5点Likert量表,3-5分定义为「可接受」:10.4秒方案74.1%可接受,标准20.8秒方案32.4%,多层CT 100%。
- 与多层CT相比,83%的10.4秒锥形束CT图像质量被认为足以支持直接至血管造影室筛选,标准20.8秒方案仅11%。
临床应用与阅片要点
- 改进10.4秒锥形束CT方案可改善脑实质图像质量,可能支持急性缺血性卒中取栓患者的直接至血管造影室流程。
- 灰白质分化改善59%、噪声降低63%、伪影减少50%,有助于在血管造影室评估脑实质。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Cone beam CT (CBCT) imaging assessment of acute ischemic stroke (AIS) patients with large-vessel occlusion (LVO) in the angiosuite may improve stroke workflow and decrease time to recanalization. In order for this workflow to gain widespread acceptance, current CBCT imaging needs further development to improve image quality. Our study aimed to compare the image quality of a new CBCT protocol performed directly in the angiosuite with imaging from multidetector CT as a gold standard.
METHODS: AIS patients with an LVO who were candidates for endovascular treatment were prospectively included in this study. Following conventional multidetector CT (MDCT), patients underwent unenhanced cone beam CT (XperCT, Philips) imaging in the angiosuite, using two different protocols: a standard 20.8 s XperCT and/or an improved 10.4 s XperCT protocol. Images were evaluated using both qualitative and quantitative methods.
RESULTS: We included 65 patients in the study. Patients received CBCT imaging prior to endovascular treatment; 18 patients were assessed with a standard 20.8 s protocol scans and 47 with a newer 10.4 s scan. The quantitative analysis showed that the mean contrast-to-noise ratio (CNR) was significantly higher for the newer 10.4 s protocol compared with the 20.8 s protocol (2.08 +/- 0.64 vs. 1.15 +/- 0.27, p < 0.004) and the mean image noise was significantly lower for the 10.4 s XperCTs when compared with the 20.8 s XperCTs (6.30 +/- 1.34 vs. 7.82 +/- 2.03, p=<0.003). Qualitative analysis, including 6 measures of image quality, demonstrated that 74.1 % of the 10.4 s XperCT scans were ranked as 'Acceptable' for assessing parenchymal imaging in AIS patients(scoring 3-5 points on a 5-point Likert-scale), compared with 32.4 % of the standard 20.8 s XperCT and 100 % of the MDCT scans. Compared to the MDCT studies, 83 % of the 10.4 s XperCT scans were deemed sufficient image quality for a direct-to-angiosuite selection, compared to only 11 % for the standard 20.8 s scans. The largest image quality improvements included grey/white matter differentiation (59 % improvement), and reduction of image noise and artefacts (63 % & 50 % improvement, respectively).
CONCLUSIONS: Continued advances in cone-beam CT allow marked improvements in image quality for the assessment of brain parenchyma, which supports a direct-to-angiosuite approach for AIS patients eligible for thrombectomy treatment.
引用
- Nicholson P, Cancelliere NM, Bracken J, et al. Novel flat-panel cone-beam CT compared to multi-detector CT for assessment of acute ischemic stroke: A prospective study. Eur J Radiol. 2021. PMID: 33725654. doi:10.1016/j.ejrad.2021.109645
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。