320排CT透视引导介入的疗效与辐射暴露首次经验

First experience of efficacy and radiation exposure in 320-detector row CT fluoroscopy-guided interventions.
PubMed 文献 · Br J Radiol · 2021-02-05

回顾性队列分析231例320排CT透视引导介入,总体技术成功率98.7%(228/231),活检临床成功率93.5%(146/156),3例主要并发症均发生于胸部活检。中位总辐射剂量522.4(393.4-819.8)mGy*cm,其中87%来自介入前后CT系列,仅11.4%来自CT透视引导介入;多元线性回归显示男性、体重指数、引流、介入时间和作为介入后CT的螺旋扫描与更高剂量显著相关。

要点

  • 总体技术成功率98.7%(228/231),活检、引流、抽吸分别为98.7%(154/156)、98.5%(66/67)、100%(8/8)。
  • 3例主要并发症均发生于胸部活检。
  • 中位总辐射剂量522.4(393.4-819.8)mGy*cm,87%来自介入前后CT系列,仅11.4%来自CT透视引导介入。

诊断标准与关键数值

  • 总体技术成功率为98.7%(228/231)。
  • 活检技术成功率为98.7%(154/156),临床成功率为93.5%(146/156)。
  • 引流技术成功率为98.5%(66/67),抽吸技术成功率为100%(8/8)。

临床应用与阅片要点

  • 中位总辐射剂量为522.4(393.4-819.8)mGy*cm。
  • 总辐射剂量中87%施加于介入前和介入后CT系列,介入后CT占总辐射剂量的24.4%,CT透视引导介入仅占11.4%。
  • 多元线性回归显示男性、体重指数、引流、介入时间和作为介入后CT的螺旋扫描与更高剂量显著相关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We investigated the efficacy and exposure to radiation in 320-detector row computed tomography fluoroscopy-guided (CTF-guided) interventions.

METHODS: We analysed 231 320-detector row CTF-guided interventions (207 patients over 2 years and 6 months) in terms of technical success rates, clinical success rates, complications, scanner settings, overall radiation doses (dose-length product, mGy*cm), patient doses of peri-interventional CT series, and interventional CT (including CTF), as a retrospective cohort study. The relationships between patient radiation dose and interventional factors were assessed using multivariate analysis.

RESULTS: Overall technical success rate was 98.7% (228/231). The technical success rates of biopsies, drainages, and aspirations were 98.7% (154/156), 98.5% (66/67), and 100% (8/8), respectively. The clinical success rate of biopsies was 93.5% (146/156). All three major complications occurred in chest biopsies. The median total radiation dose was 522.4 (393.4-819.8) mGy*cm. Of the total radiation dose, 87% was applied during the pre- and post-interventional CT series. Post-interventional CT accounted for 24.4% of the total radiation dose. Only 11.4% of the dose was applied by CTF-guided intervention. Multilinear regression demonstrated that male sex, body mass index, drainage, intervention time, and helical scan as post-interventional CT were significantly associated with higher dose.

CONCLUSION: The 320-detector row CTF interventions achieved a high success rate. Dose reduction in post-interventional CT provides patient dose reduction without decreasing the technical success rates.

ADVANCES IN KNOWLEDGE: This is the first study on the relationship between various interventional outcomes and patient exposure to radiation in 320-detector row CTF-guided interventions, suggesting a new perspective on dose reduction.