放射性栓塞模拟治疗中肝脏CT容积测量的观察者内与观察者间重复性
该回顾性研究纳入2019年5月至2020年2月间17例连续接受TARE治疗前模拟评估的患者,比较介入放射科医师与三名放射技师在MDCT上测量健康肝体积、全肝实质体积及全肝体积的重复性。结果显示放射技师观察者内信度0.997至1.000(95%CI),观察者间重现性0.965至0.999(95%CI),准确度0.964至0.999(95%CI)。作者认为经过适当培训的放射技师在评估TARE计划所需肝脏CTV数据方面可能具有与介入放射科医师相同的准确度。
要点
- 四名观察者检出的每例患者肝脏病灶数目一致。
- 放射技师观察者内信度0.997至1.000(95%CI),观察者间重现性0.965至0.999(95%CI)。
- 放射技师评估准确度为0.964至0.999(95%CI)。
关键影像表现/征象
- 四名观察者检出的每例患者肝脏病灶数目一致。
- 评估参数包括健康肝体积(HLV)、TARE涉及的全肝实质体积(THV=健康肝脏与TTV=肿瘤)及全肝体积(WLV)。
诊断标准与关键数值
- 放射技师观察者内信度为0.997至1.000(95%CI)。
- 放射技师间观察者间重现性为0.965至0.999(95%CI)。
- 放射技师评估准确度为0.964至0.999(95%CI)。
临床应用与阅片要点
- 作者认为经过适当培训的放射技师在评估TARE计划所需肝脏CTV数据方面可能具有与介入放射科医师相同的准确度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The careful evaluation of MDCT is an essential step for the treatment planning in pre-treatment imaging work-up for Trans-Arterial Radio Embolization (TARE). It may provide unique volumetric data (CTVs), which are information useful for an effective and safe TARE. The purpose of this study is to demonstrate that the radiographer is able to calculate CTVs of TARE simulation with the same precision as the interventional radiologist.
METHODS: This study retrospectively considers 17 consecutive patients (8 males, 9 females; mean age 66.3 ± 13.2 years) who underwent pre-treatment work-up for TARE, between May 2019 and February 2020 (trial ID:2234 - protocol). For each patient, four specific parameters are evaluated from MDCT achieved during treatment simulation: healthy liver volume (HLV), the whole hepatic parenchyma (THV = healthy liver and TTV = tumour) involved by TARE, and whole liver volume (WLV). Four independent observers-R1 (expert interventional radiologist), T1, T2, and T3 (radiographers, with different experiences in the field of interventional radiology)-are involved in the imaging analysed.
RESULTS: All the 4 observers detected the same number of hepatic lesion(s) per patient. Regarding the three radiographers, the intra-observer reliability for CTVs is very high 0.997 to 1.000 (95%CI). Also inter-observer reproducibility between radiographers is excellent regarding CTVs, 0.965 to 0.999 (95%CI). The accuracy of radiographer evaluation is very high 0.964 to 0.999 (95%CI).
CONCLUSIONS AND IMPLICATIONS FOR PRACTICE: The high intra- and inter-observer reproducibility shows that a properly trained radiographers might have the same accuracy as interventional radiologists, in assessing liver CTV data for planning TARE.
引用
- Masperi A, Cubadda V, Bombelli L, et al. Intra- inter-observer repeatability in liver computed tomography volumetry in patients undergoing radioembolization simulation. Abdom Radiol (NY). 2021. PMID: 33630128. doi:10.1007/s00261-021-02980-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。