文献速递腹部MR

肝脏MRI质子密度脂肪分数量化中不同感兴趣区采样策略的重复性与准确性

Repeatability and accuracy of various region-of-interest sampling strategies for hepatic MRI proton density fat fraction quantification.
PubMed 文献 · Abdom Radiol (NY) · 2021-02-20

本研究为二次分析,纳入29例已知或疑似非酒精性脂肪性肝病受试者,每例完成三次检查、每次三次采集(共九次采集),评估511种ROI采样策略的重复性与误差。结果显示ROI越多重复性越好、误差越小;60种平衡4-ROI策略均达到检查间与检查内ICC>0.998,MAE均<1%。

要点

  • 平衡4-ROI策略检查间与检查内ICC均>0.998,MAE均<1%。
  • ROI数量增加时ICC升高、重复性系数降低、平均绝对误差降低。
  • 平衡采样策略整体上ICC更高、重复性系数更低、MAE更低(P<0.001)。

临床应用与阅片要点

  • 作者结论认为平衡4-ROI策略提供高重复性和低误差,原文未据此给出具体阅片操作推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate repeatability of ROI-sampling strategies for quantifying hepatic proton density fat fraction (PDFF) and to assess error relative to the 9-ROI PDFF.

METHODS: This was a secondary analysis in subjects with known or suspected nonalcoholic fatty liver disease who underwent MRI for magnitude-based hepatic PDFF quantification. Each subject underwent three exams, each including three acquisitions (nine acquisitions total). An ROI was placed in each hepatic segment on the first acquisition of the first exam and propagated to other acquisitions. PDFF was calculated for each of 511 sampling strategies using every combination of 1, 2, …, all 9 ROIs. Intra- and inter-exam intra-class correlation coefficients (ICCs) and repeatability coefficients (RCs) were estimated for each sampling strategy. Mean absolute error (MAE) was estimated relative to the 9-ROI PDFF. Strategies that sampled both lobes evenly ("balanced") were compared with those that did not ("unbalanced") using two-sample t tests.

RESULTS: The 29 enrolled subjects (23 male, mean age 24 years) had mean 9-ROI PDFF 11.8% (1.1-36.3%). With more ROIs, ICCs increased, RCs decreased, and MAE decreased. Of the 60 balanced strategies with 4 ROIs, all (100%) achieved inter- and intra-exam ICCs > 0.998, 55 (92%) achieved intra-exam RC < 1%, 50 (83%) achieved inter-exam RC < 1%, and all (100%) achieved MAE < 1%. Balanced sampling strategies had higher ICCs and lower RCs, and lower MAEs than unbalanced strategies in aggregate (p < 0.001 for comparisons between balanced vs. unbalanced strategies).

CONCLUSION: Repeatability improves and error diminishes with more ROIs. Balanced 4-ROI strategies provide high repeatability and low error.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Hong CW, Cui JY, Batakis D, et al. Repeatability and accuracy of various region-of-interest sampling strategies for hepatic MRI proton density fat fraction quantification. Abdom Radiol (NY). 2021. PMID: 33609166. doi:10.1007/s00261-021-02965-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。