深度学习图像重建低剂量全身CT:图像质量与病灶检出

Low-dose whole-body CT using deep learning image reconstruction: image quality and lesion detection.
PubMed 文献 · Br J Radiol · 2021-02-22

本研究对59例恶性肿瘤监测患者行低剂量门静脉期全身CT,分别用混合迭代重建(LD-IR)和深度学习图像重建(LD-DLIR),并与一年内标准剂量混合迭代重建(SD-IR)比较。LD-DLIR的图像质量和病灶检出率与SD-IR相当,背景噪声更低、SNR更高,CTDIvol和DLP显著低于SD。结论认为DLIR可在维持图像质量和病灶检出率的同时降低辐射剂量>75%。

要点

  • LD-DLIR图像质量与病灶检出率与标准剂量SD-IR相当。
  • LD-DLIR背景噪声显著低于SD-IR和LD-IR,SNR显著提高。
  • 低剂量扫描CTDIvol 2.9 mGy、DLP 216.2 mGy•cm,显著低于标准剂量。
英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate image quality and lesion detection capabilities of low-dose (LD) portal venous phase whole-body computed tomography (CT) using deep learning image reconstruction (DLIR).

METHODS: The study cohort of 59 consecutive patients (mean age, 67.2 years) who underwent whole-body LD CT and a prior standard-dose (SD) CT reconstructed with hybrid iterative reconstruction (SD-IR) within one year for surveillance of malignancy were assessed. The LD CT images were reconstructed with hybrid iterative reconstruction of 40% (LD-IR) and DLIR (LD-DLIR). The radiologists independently evaluated image quality (5-point scale) and lesion detection. Attenuation values in Hounsfield units (HU) of the liver, pancreas, spleen, abdominal aorta, and portal vein; the background noise and signal-to-noise ratio (SNR) of the liver, pancreas, and spleen were calculated. Qualitative and quantitative parameters were compared between the SD-IR, LD-IR, and LD-DLIR images. The CT dose-index volumes (CTDIvol) and dose-length product (DLP) were compared between SD and LD scans.

RESULTS: The image quality and lesion detection rate of the LD-DLIR was comparable to the SD-IR. The image quality was significantly better in SD-IR than in LD-IR (p < 0.017). The attenuation values of all anatomical structures were comparable between the SD-IR and LD-DLIR (p = 0.28-0.96). However, background noise was significantly lower in the LD-DLIR (p < 0.001) and resulted in improved SNRs (p < 0.001) compared to the SD-IR and LD-IR images. The mean CTDIvol and DLP were significantly lower in the LD (2.9 mGy and 216.2 mGy•cm) than in the SD (13.5 mGy and 1011.6 mGy•cm) (p < 0.0001).

CONCLUSION: LD CT images reconstructed with DLIR enable radiation dose reduction of >75% while maintaining image quality and lesion detection rate and superior SNR in comparison to SD-IR.

ADVANCES IN KNOWLEDGE: Deep learning image reconstruction algorithm enables around 80% reduction in radiation dose while maintaining the image quality and lesion detection compared to standard-dose whole-body CT.