单次激发压缩感知电影成像用于左心室功能与应变分析的可行性
研究纳入37名受试者,对比单次激发压缩感知电影序列与常规分段电影序列在心脏磁共振中定量左心室容积与应变的能力。压缩感知电影测量的左心室容积和心肌质量与分段电影强相关(ICC>0.798),但整体应变与达峰应变时间标准差的相关性差至中等(ICC 0.045–0.706),且整体应变值被低估,两种技术的应变值不可互换。
要点
- 单次激发压缩感知电影测量左心室容积和心肌质量与常规分段电影强相关(ICC>0.798)。
- 压缩感知电影的整体应变与达峰应变时间标准差和分段电影相关性差至中等(ICC 0.045–0.706)。
- 压缩感知电影的整体应变值均低于分段电影,两种技术的应变值不可互换。
关键影像表现/征象
- 单次激发压缩感知电影测量的左心室容积和心肌质量与分段电影强相关(ICC>0.798)。
- 单次激发压缩感知电影存在轻度收缩末期容积高估,导致射血分数低估。
- 单次激发压缩感知电影得出的整体应变和达峰应变时间标准差与分段电影相关性差至中等(ICC 0.045–0.706)。
- 单次激发压缩感知电影得出的所有整体应变值均低于分段电影得出的值,但两种电影在径向和纵向达峰应变时间标准差上无显著差异。
- 在患者相关因素中,心率是压缩感知电影整体纵向应变低估的强预测因素(p=0.039)。
诊断标准与关键数值
- 单次激发压缩感知电影与分段电影左心室容积和心肌质量的组内相关系数ICC>0.798。
- 单次激发压缩感知电影与分段电影整体应变和达峰应变时间标准差的组内相关系数ICC为0.045–0.706。
- 压缩感知电影整体纵向应变低估与心率相关(p=0.039)。
- 压缩感知电影和分段电影得出的左心室应变观察者间与观察者内变异性均为好至优。
临床应用与阅片要点
- 单次激发压缩感知电影可用于左心室容积和心肌质量的定量评估,与常规分段电影强相关(ICC>0.798)。
- 单次激发压缩感知电影的整体应变值低于分段电影,且两种技术的应变值不可互换,解读时应变结果时需注意所用序列。
- 心率是压缩感知电影整体纵向应变低估的强预测因素(p=0.039),阅片时需结合心率考量应变低估的可能。
局限性
- 单次激发压缩感知电影与分段电影的整体应变和达峰应变时间标准差相关性差至中等(ICC 0.045–0.706),两种技术的应变值不可互换。
- 单次激发压缩感知电影存在轻度收缩末期容积高估,导致射血分数低估。
- 单次激发压缩感知电影得出的所有整体应变值均低于分段电影得出的值。
- 心率是压缩感知电影整体纵向应变低估的强预测因素(p=0.039)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To compare single-shot compressed sensing (CS) cine imaging with conventional segmented cine imaging for reliable quantification of left ventricular (LV) volume and strain assessment during cardiac magnetic resonance imaging (CMRI).
MATERIALS AND METHODS: Thirty-seven participants underwent both single-shot CS and conventional segmented cines that covered the entire LV. LV volumetric and strain values were obtained. LV volumes, global strain, the standard deviation of time to peak strain (SD-TPS) in the radial, longitudinal, and circumferential directions were compared using the Student's t-test and intraclass correlation coefficient (ICC). Interobserver and intra-observer variabilities of the LV strain values for the two cines method were determined using ICC.
RESULTS: Single-shot CS cine-derived LV volumes and myocardial mass measurements correlated strongly with segmented cines (ICC >0.798) and minor systematic end-systolic volume overestimations resulting in ejection fraction underestimations. Single-shot CS cine-derived global strain and SD-TPS were poorly to moderately correlated with segmented cines (ICC from 0.045-0.706). All global strain values derived from single-shot CS cines were underestimated compared with segmented cine-derived values; however, no significant differences in radial and longitudinal SD-TPS between the two cines were found. Among the patient-related factors, heart rate was a strong predictive factor of global longitudinal strain underestimations (p=0.039) in the CS cines. Inter- and intra-observer LV strain variabilities derived from CS and segmented cines were good to excellent.
CONCLUSION: Single-shot CS cine CMRI is feasible for the quantitative assessment of LV function. Currently, strain values derived from the two techniques are not interchangeable.