自导航与导航门控3D MRI序列在TAVI术前无增强主动脉根部测量中的前瞻性比较
前瞻性比较原型自导航与导航门控非增强3D全心MRA序列在TAVI术前主动脉根部测量中的图像质量、可靠性和瓣膜尺寸选择,并以CTA为参照。27例患者(83±5岁,41%男性)行两种MRA,其中15例(56%)加做CTA。自导航MRA采集时间更短(5.5±1 min vs 6.5±2 min,p=0.029),主动脉根部测量观察者间相关性高至非常高,自导航MRA与CTA的瓣膜尺寸选择等效(κ=1),但4例(15%)自导航MRA上各有一支冠状动脉开口显示不清。
要点
- 自导航MRA采集时间较导航门控MRA缩短(5.5±1 min vs 6.5±2 min,p=0.029)。
- 自导航MRA与导航门控MRA的SNR和CNR无显著差异(p=0.471和0.445)。
- 自导航MRA与CTA的理论瓣膜尺寸选择等效(κ=1),但4例(15%)有一支冠状动脉开口显示不清。
诊断标准与关键数值
- 主观图像质量采用4点Likert量表评分,自导航MRA中位评分1.5(IQR 1-3),导航门控MRA中位评分1(IQR 1-2),p=0.059。
- 自导航MRA与导航门控MRA的SNR和CNR无显著差异(p=0.471和0.445)。
- 自导航MRA与导航门控MRA的主动脉根部测量观察者间相关性分别为r=0.75至0.94和r=0.85至0.96(均p<0.0001)。
- 自导航MRA与CTA的理论人工瓣膜尺寸选择等效(κ=1)。
临床应用与阅片要点
- 自导航MRA可用于TAVI术前主动脉根部测量和瓣膜尺寸选择,采集时间较导航门控MRA缩短。
- 自导航MRA上4例(15%)各有一支冠状动脉开口显示不清(右冠状动脉3例,左主干1例),阅片时需注意评估冠状动脉开口。
局限性
- 自导航MRA上4例(15%)各有一支冠状动脉开口显示不清(右冠状动脉3例,左主干1例)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To prospectively compare image-quality, reliability and graft sizing of a prototype self-navigated and a navigator-gated non-contrast three dimensional (3D) whole-heart magnetic-resonance-angiography (MRA) sequence with computed-tomography-angiography (CTA) for planning transcatheter-aortic-valve-implantation (TAVI).
METHODS: Self- and navigator-gated 1.5 T MRA were performed in 27 patients (aged 83 ± 5 years, 41 % male) for aortic root sizing and coronary ostia height measurements; 15 (56 %) patients underwent additional CTA. Subjective-image quality was graded on a 4-point Likert scale, objective MRA image-quality was assessed by signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Continuous MRA and CTA measurements were analyzed with regression and Bland-Altman analysis, valve sizing by kappa statistics.
RESULTS: Median image-quality as rated by two observers was 1.5 [interquartile range (IQR) 1-3] for self-navigated MRA and 1 [IQR 1-2] for navigator-gated MRA (p = 0.059). SNR and CNR were comparable between MRA sequences (p = 0.471 and 0.445, respectively). Acquisition time was shorter for self-navigated MRA compared to navigator-gated MRA (5.5 ± 1 min vs, 6.5 ± 2 min, p = 0.029). Inter-observer correlation of aortic root measurements was high to very high for both self- and navigator-gated MRA (r = 0.75 to 0.94 and r = 0.85 to 0.96, respectively, all p < 0.0001). Theoretical prosthetic valve sizing of self-navigated MRA and CTA was equivalent (κ = 1). However, in four patients (15 %) one coronary ostium each (right coronary artery 3, left main artery 1) was not clearly definable on self-navigated MRA.
CONCLUSION: Self-navigated MRA enables aortic annulus TAVI measurements without significant difference to navigator-gated MRA at shortened acquisition time. Prosthesis sizing by self-navigated MRA measurements is equivalent to navigator-gated MRA and CTA-based choice.
引用
- Pamminger M, Kranewitter C, Kremser C, et al. Self-navigated versus navigator-gated 3D MRI sequence for non-enhanced aortic root measurement in transcatheter aortic valve implantation. Eur J Radiol. 2021. PMID: 33578090. doi:10.1016/j.ejrad.2021.109573
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。