简易下体负压装置用于MRI评估心血管功能变化的可行性研究

The design and use of a simple device for the MRI assessment of changes in cardiovascular function by lower-body negative-pressure-simulated reduction of central blood volume.
PubMed 文献 · Clin Radiol · 2021-02-23

研究使用本地设计的简易下体负压(LBNP)装置和1.5T MRI,在14名健康志愿者中于0、-5、-10、-20 mmHg压力水平前瞻性评估前负荷降低对心室容积和大血管血流的影响。结果显示-5 mmHg时右心室输出量和肺动脉净流量下降,-20 mmHg时主动脉和肺动脉净流量及左右心室舒张末期容积均下降。作者认为MRI兼容LBNP装置在同一实验中测量心室容积和大动脉血流变化是可行的,可能促进更广泛心血管病理的前负荷降低研究。

要点

  • 在-5 mmHg时右心室输出量(p=0.013)和肺动脉净流量(p=0.026)较基线下降,主动脉血流和左心室容积无变化。
  • 在-20 mmHg时主动脉和肺动脉净流量(p<0.001)、左右心室舒张末期容积(p<0.001)及左右心室收缩末期容积(左p=0.038,右p=0.003)均下降。
  • 使用MRI兼容LBNP装置在同一实验中测量心室容积和大动脉血流变化是可行的。

关键影像表现/征象

  • 在-5 mmHg时,主动脉血流和左心室容积较基线无变化。
  • 在-5 mmHg时,右心室输出量(p=0.013)和肺动脉净流量(p=0.026)较基线下降。
  • 在-20 mmHg时,主动脉和肺动脉净流量(p<0.001)较基线下降。
  • 在-20 mmHg时,左和右心室舒张末期容积(p<0.001)及左和右心室收缩末期容积(左p=0.038,右p=0.003)较基线下降。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To use a locally designed and simple lower-body negative-pressure (LBNP) device and 1.5 T magnetic resonance imaging (MRI) to demonstrate the ability to assess changes in cardiovascular function during preload reduction. These effects were evaluated on ventricular volumes and great vessel flow in healthy volunteers, for which there are limited published data.

MATERIAL AND METHODS: After ethical review, 14 volunteers (mean age 33.9 ± 7 years, mean body mass index [BMI] 23.1 ± 2.5) underwent LBNP prospectively at 0, -5, -10, and -20 mmHg pressure, using a locally designed LBNP box. Expiratory breath-hold biventricular volumes, and free-breathing flow imaging of the ascending aorta and main pulmonary artery were acquired at each level of LBNP.

RESULTS: At -5 mmHg, there was no change in aortic flow or left ventricular volumes versus baseline. Right ventricular output (p=0.013) and pulmonary net flow (p=0.026) decreased. At -20 mmHg, aortic and pulmonary net flow (p<0.001) decreased, as were left and right ventricular end diastolic volume (p<0.001) and left and right end systolic volumes (p=0.038 and p=0.003 respectively).

CONCLUSIONS: Use of a MRI-compatible LBNP device is feasible to measure changes in ventricular volume and great arterial flow in the same experiment. This may enhance further research into the effects of preload reduction by MRI in a wide range of important cardiovascular pathologies.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Hamilton MCK, Neumann S, Szantho G, et al. The design and use of a simple device for the MRI assessment of changes in cardiovascular function by lower-body negative-pressure-simulated reduction of central blood volume. Clin Radiol. 2021. PMID: 33637308. doi:10.1016/j.crad.2021.01.016

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。