多中心4D血流CMR自动瓣膜追踪定量瓣膜血流的一致性评估
研究在7个中心纳入64例患者和76名健康志愿者,评估4D血流CMR自动回顾性瓣膜追踪定量瓣膜血流的观察者间一致性和瓣膜间变异。结果显示各中心和瓣膜净前向体积ICC为0.85至0.96,反流分数ICC为0.53至0.97,瓣膜间变异≤10.5%,每个瓣膜2幅电影图像较1幅预测净前向体积变异降低。
要点
- 自动回顾性瓣膜追踪定量瓣膜血流在不同中心和设备间具有强至优秀的一致性。
- 各中心瓣膜间净前向体积变异均≤10.5%,支持结果可重复。
- 每个瓣膜使用2幅电影图像较1幅可降低4个瓣膜间净前向体积变异。
诊断标准与关键数值
- 观察者间分析显示所有中心和瓣膜的净前向体积ICC为0.85至0.96,反流分数ICC为0.53至0.97。
- 所有观察者分析中瓣膜间变异低(≤10.5%)。
- 每个瓣膜2幅电影图像用于瓣膜追踪较1幅电影图像预测4个瓣膜间净前向体积变异降低(beta=-1.3;p=0.01)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study determined: 1) the interobserver agreement; 2) valvular flow variation; and 3) which variables independently predicted the variation of valvular flow quantification from 4-dimensional (4D) flow cardiac magnetic resonance (CMR) with automated retrospective valve tracking at multiple sites.
BACKGROUND: Automated retrospective valve tracking in 4D flow CMR allows consistent assessment of valvular flow through all intracardiac valves. However, due to the variance of CMR scanners and protocols, it remains uncertain if the published consistency holds for other clinical centers.
METHODS: Seven sites each retrospectively or prospectively selected 20 subjects who underwent whole heart 4D flow CMR (64 patients and 76 healthy volunteers; aged 32 years [range 24 to 48 years], 47% men, from 2014 to 2020), which was acquired with locally used CMR scanners (scanners from 3 vendors; 2 1.5-T and 5 3-T scanners) and protocols. Automated retrospective valve tracking was locally performed at each site to quantify the valvular flow and repeated by 1 central site. Interobserver agreement was evaluated with intraclass correlation coefficients (ICCs). Net forward volume (NFV) consistency among the valves was evaluated by calculating the intervalvular variation. Multiple regression analysis was performed to assess the predicting effect of local CMR scanners and protocols on the intervalvular inconsistency.
RESULTS: The interobserver analysis demonstrated strong-to-excellent agreement for NFV (ICC: 0.85 to 0.96) and moderate-to-excellent agreement for regurgitation fraction (ICC: 0.53 to 0.97) for all sites and valves. In addition, all observers established a low intervalvular variation (≤10.5%) in their analysis. The availability of 2 cine images per valve for valve tracking compared with 1 cine image predicted a decreasing variation in NFV among the 4 valves (beta = -1.3; p = 0.01).
CONCLUSIONS: Independently of locally used CMR scanners and protocols, valvular flow quantification can be performed consistently with automated retrospective valve tracking in 4D flow CMR.