文献速递心血管儿科影像MR

压缩感知技术减少儿童及青年心脏电影平衡稳态自由进动磁共振成像的扫描时间与屏气次数

Use of compressed sensing to reduce scan time and breath-holding for cardiac cine balanced steady-state free precession magnetic resonance imaging in children and young adults.
PubMed 文献 · Pediatr Radiol · 2021-02-10

在28例儿童及青年患者中,于3T磁共振采用回顾性心电门控电影平衡稳态自由进动序列,对比常规并行成像与厂商优化的并行成像/压缩感知方案。压缩感知使总扫描时间从147±33秒降至83±28秒(降低43%),屏气次数从10±4次降至5±3次(降低50%),图像定性评分及双心室容积与功能定量分析均无显著差异。

要点

  • 压缩感知电影序列每层扫描7±2秒降至4±2秒,总扫描时间147±33秒降至83±28秒。
  • 屏气次数由常规10±4次减少至压缩感知5±3次,降幅约50%。
  • 血-心肌对比、边缘定义及伪影的定性评分与双心室容积功能定量均无显著差异。
英文原文(摘要)· 点击展开

BACKGROUND: Conventional pediatric volumetric MRI acquisitions of a short-axis stack typically require multiple breath-holds under anesthesia.

OBJECTIVE: Here, we aimed to validate a vendor-optimized compressed-sensing approach to reduce scan time during short-axis balanced steady-state free precession (bSSFP) cine imaging.

MATERIALS AND METHODS: Imaging was performed in 28 patients (16±9 years) in this study on a commercial 3-tesla (T) scanner using retrospective electrocardiogram-gated cine bSSFP. Cine short-axis images covering both ventricles were acquired with conventional parallel imaging and a vendor-optimized parallel imaging/compressed-sensing approach. Qualitative Likert scoring for blood-myocardial contrast, edge definition, and presence of artifact was performed by two experienced radiologists. Quantitative comparisons were performed including biventricular size and function. A paired t-test was used to detect significant differences (P<0.05).

RESULTS: Scan duration was 7±2 s/slice for conventional imaging (147±33 s total) vs. 4±2 s/slice for compressed sensing (83±28 s total). No significant differences were found with qualitative image scores for blood-myocardial contrast, edge definition, and presence of artifact. No significant differences were found in volumetric analysis between the two sequences. The number of breath-holds was 10±4 for conventional imaging and 5±3 for compressed sensing.

CONCLUSION: Compressed sensing allowed for a 50% reduction in the number of breath-holds and a 43% reduction in the total scan time without differences in the qualitative or quantitative measurements as compared to the conventional technique.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Naresh NK, Malone L, Fujiwara T, et al. Use of compressed sensing to reduce scan time and breath-holding for cardiac cine balanced steady-state free precession magnetic resonance imaging in children and young adults. Pediatr Radiol. 2021. PMID: 33566124. doi:10.1007/s00247-020-04952-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。