上腹部增强CT深度学习重建:与迭代重建直接比较图像质量相似且辐射剂量更低

Deep learning reconstruction for contrast-enhanced CT of the upper abdomen: similar image quality with lower radiation dose in direct comparison with iterative reconstruction.
PubMed 文献 · Eur Radiol · 2021-02-08

连续纳入100例同时行胸部及腹部增强CT的患者,比较深度学习重建(DLIR)胸CT、自适应统计迭代重建(ASiR)胸CT与ASiR腹部CT在上腹部的图像质量。DLIR胸CT辐射剂量不足腹部CT的50%,客观噪声最低、SNR与CNR最高,主观上ASiR腹部CT噪声评分与整体图像质量更优,但三名阅片者中两名更常偏好DLIR。

要点

  • DLIR胸CT上腹部图像噪声8.01±2.81,低于ASiR胸CT的14.8±2.56与ASiR腹部CT的12.3±2.28(p<.001)。
  • 主观噪声ASiR腹部CT评分2.94±0.23优于DLIR的2.87±0.26(p<.001),DLIR空间分辨率更优(2.60±0.34 vs. 2.44±0.31,p=.02)。
  • DLIR重建图像上重建算法所致变形伪影更常见,但仅0.3%的病例报告诊断困难。

核心结论

  • 在不足50%辐射剂量下,深度学习重建胸CT在上腹部显示与专用腹部CT相当的图像质量,且被多数阅片者偏好。

研究设计与样本

  • 连续纳入100例同时行胸部及腹部增强CT的患者,采集三组图像:ASiR胸CT(50% blending)、DLIR胸CT、ASiR腹部CT(40% blending)。
  • 辐射剂量按每次扫描优化,平均CTDIvol为胸部CT 3.19±1.53 mGy、腹部CT 7.10±1.88 mGy。
  • 提取覆盖上腹部的图像,测量图像噪声、信噪比(SNR)与对比噪声比(CNR),并由三名放射科医师独立主观评估噪声、空间分辨率、伪影及整体图像质量,另选择每例最偏好的重建技术。

关键影像表现/征象

  • DLIR胸CT、ASiR胸CT与ASiR腹部CT的平均测量噪声分别为8.01±2.81、14.8±2.56、12.3±2.28(p<.001)。
  • DLIR的SNR与CNR均最佳(p<.001)。
  • 主观分析中ASiR腹部CT主观噪声评分2.94±0.23优于DLIR的2.87±0.26(p<.001),DLIR空间分辨率评分2.60±0.34优于ASiR腹部CT的2.44±0.31(p=.02)。
  • ASiR腹部CT整体图像质量更优(p=.001),但三名阅片者中两名更常偏好DLIR。

局限性

  • 重建算法所致变形伪影在DLIR重建图像上更常见,但仅0.3%的病例报告诊断困难。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the effect of a commercial deep learning algorithm on the image quality of chest CT, focusing on the upper abdomen.

METHODS: One hundred consecutive patients who simultaneously underwent contrast-enhanced chest and abdominal CT were collected. The radiation dose was optimized for each scan (mean CTDIvol: chest CT, 3.19 ± 1.53 mGy; abdominal CT, 7.10 ± 1.88 mGy). Three image sets were collected: chest CT reconstructed with an adaptive statistical iterative reconstruction (ASiR-CHT; 50% blending), chest CT with a deep learning algorithm (DLIR-CHT), and abdominal CT with ASiR (ASiR-ABD; 40% blending). Afterwards, the images covering the upper abdomen were extracted, and image noise, the signal-to-noise ratio (SNR), and the contrast-to-noise ratio (CNR) were measured. For subjective evaluation, three radiologists independently assessed noise, spatial resolution, presence of artifacts, and overall image quality. Additionally, readers selected the most preferable reconstruction technique among three image sets for each case.

RESULTS: The average measured noise for DLIR-CHT, ASiR-CHT, and ASiR-ABD was 8.01 ± 2.81, 14.8 ± 2.56, and 12.3 ± 2.28, respectively (p < .001). Deep learning-based image reconstruction (DLIR) also showed the best SNR and CNR (p < .001). However, in the subjective analysis, ASiR-ABD showed less subjective noise than DLIR (2.94 ± 0.23 vs. 2.87 ± 0.26; p < .001), while DLIR showed better spatial resolution (2.60 ± 0.34 vs. 2.44 ± 0.31; p = .02). ASiR-ABD showed a better overall image quality (p = .001), but two of the three readers preferred DLIR more frequently.

CONCLUSION: With < 50% of the radiation dose, DLIR chest CT showed comparable image quality in the upper abdomen to that of dedicated abdominal CT and was preferred by most readers.

KEY POINTS: • With < 50% radiation dose, a deep learning algorithm applied to contrast-enhanced chest CT exhibited better image noise and signal-to-noise ratio than standard abdominal CT with the ASiR technique. • Pooled readers mostly preferred deep learning algorithm-reconstructed contrast-enhanced chest CT reconstructed using a standard ASiR-reconstructed abdominal CT. • Reconstruction algorithm-induced distortion artifacts were more frequently observed on deep learning algorithm-reconstructed images, but diagnostic difficulty was reported in only 0.3% of cases.