文献速递神经技术与安全MR

颅内脑电图电极在1.5 T和3 T MRI中的临床安全性:单中心经验与文献回顾

Clinical safety of intracranial EEG electrodes in MRI at 1.5 T and 3 T: a single-center experience and literature review.
PubMed 文献 · Neuroradiology · 2021-02-04

回顾165例连续颅内脑电图监测入院患者,共135例患者140次入院完成184次MRI扫描,包括118次1.5 T结构MRI和66次3 T功能MRI,方案避免使用高比吸收率序列。1.5 T和3 T扫描期间及之后均未见不良事件或并发症(除电极植入相关者外),也无临床或影像证据显示原有植入相关并发症加重。

要点

  • 颅内EEG电极患者行1.5 T或3 T MRI时,避免高比吸收率序列,未见临床或影像并发症。
  • MRI后原有植入相关并发症无临床或影像加重证据。
  • 扫描安全性评价涵盖165例连续入院中的135例患者140次入院184次MRI。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Intracranial electroencephalography (EEG) can be a critical part of presurgical evaluation for drug resistant epilepsy. With the increasing use of intracranial EEG, the safety of these electrodes in the magnetic resonance imaging (MRI) environment remains a concern, particularly at higher field strengths. However, no studies have reported the MRI safety experience of intracranial electrodes at 3 T. We report an MRI safety review of patients with intracranial electrodes at 1.5 and 3 T.

METHODS: One hundred and sixty-five consecutive admissions for intracranial EEG monitoring were reviewed. A total of 184 MRI scans were performed on 135 patients over 140 admissions. These included 118 structural MRI studies at 1.5 T and 66 functional MRI studies at 3 T. The magnetic resonance (MR) protocols avoided the use of high specific energy absorption rate sequences that could result in electrode heating. The intracranial implantations included 114 depth, 15 subdural, and 11 combined subdural and depth electrodes. Medical records were reviewed for patient-reported complications and radiologic complications related to these studies. Pre-implantation, post-implantation, and post-explantation imaging studies were reviewed for potential complications.

RESULTS: No adverse events or complications were seen during or after MRI scanning at 1.5 or 3 T apart from those attributed to electrode implantation. There was also no clinical or imaging evidence of worsening of pre-existing implantation-related complications after MR imaging.

CONCLUSION: No clinical or radiographic complications are seen when performing MRI scans at 1.5 or 3 T on patients with implanted intracranial EEG electrodes while avoiding high specific energy absorption rate sequences.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Peedicail JS, Poulin T, Scott JN, et al. Clinical safety of intracranial EEG electrodes in MRI at 1.5 T and 3 T: a single-center experience and literature review. Neuroradiology. 2021. PMID: 33543360. doi:10.1007/s00234-021-02661-7

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。