Pipeline栓塞装置治疗破裂颅内动脉瘤:系统评价与个体患者数据的汇总分析
该系统评价纳入2011年1月至2020年3月MEDLINE、EMBASE和Scopus数据库中12项研究共145例患者145个破裂颅内动脉瘤,评估Pipeline栓塞装置(PED)的安全性和有效性。动脉瘤平均大小5.9 mm,多数为血泡样(51.0%)或夹层(26.9%)动脉瘤。PED置入后再破裂率为2.1%,影像学随访患者中完全闭塞率为87.5%,症状性神经并发症发生率为16.5%。
要点
- 破裂颅内动脉瘤经PED治疗后再破裂率为2.1%,影像学随访完全闭塞率为87.5%。
- 动脉瘤平均大小5.9 mm,血泡样(51.0%)和夹层(26.9%)形态占多数。
- 症状性神经并发症发生率为16.5%,较大动脉瘤大小是再破裂的显著预测因素(P = .008)。
诊断标准与关键数值
- 影像学随访患者中完全闭塞率为87.5%。
- PED置入后再破裂率为2.1%(3/145)。
- 症状性神经并发症发生率为16.5%。
- 单因素logistic回归显示较大动脉瘤大小是动脉瘤再破裂的显著预测因素(P = .008)。
局限性
- 分析受纳入数据质量限制,多数来自小型病例系列,代表III级医学证据。
- 无研究以盲法或独立裁定方式评估结局。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: The Pipeline Embolization Device (PED) is a flow-diverting stent for the treatment of intracranial aneurysms and is used off-label for a subset of ruptured aneurysms not amenable to traditional treatment.
PURPOSE: Our aim was to evaluate the safety and efficacy of the PED for treatment of ruptured intracranial aneurysms.
DATA SOURCES: A systematic review of the MEDLINE, EMBASE, and Scopus data bases from January 2011 to March 2020 was performed for articles reporting treatment of ruptured intracranial aneurysms with the PED.
STUDY SELECTION: A total of 12 studies comprising 145 patients with 145 treated aneurysms were included for analysis.
DATA ANALYSIS: Individual patient data were collected. Nonparametric tests were used to compare differences among patients. Logistic regression was used to determine an association with outcome variables.
DATA SYNTHESIS: Mean aneurysm size was 5.9 mm, and most were blister (51.0%) or dissecting (26.9%) in morphology. Three (2.1%) aneurysms reruptured following PED placement. Univariate logistic regression identified larger aneurysm size as a significant predictor of aneurysm rerupture (P = .008). Of patients with radiographic follow-up, 87.5% had complete aneurysm occlusion. Symptomatic neurologic complications occurred in 16.5%.
LIMITATIONS: Analysis was limited by the quality of the included data, most of which were from small case series representing class III medical evidence. No study assessed outcome in a blinded or independently adjudicated manner.
CONCLUSIONS: Most ruptured aneurysms treated with the PED were blister or dissecting aneurysms. Treatment was associated with a rerupture rate of 2.1% and a complete occlusion rate of 87.5%.
引用
- Foreman PM, Ilyas A, Cress MC, et al. Ruptured Intracranial Aneurysms Treated with the Pipeline Embolization Device: A Systematic Review and Pooled Analysis of Individual Patient Data. AJNR Am J Neuroradiol. 2021. PMID: 33602746. doi:10.3174/ajnr.A7002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。