超低剂量CT在桡骨远端骨折三维打印中的应用

Application of ultra-low-dose CT in 3D printing of distal radial fractures.
PubMed 文献 · Eur J Radiol · 2020-12-29

本研究纳入76例桡骨远端骨折患者,随机分为标准剂量组(120 kV、100 mA,38例)和超低剂量组(80 kV、10 mA,38例),比较320排CT图像质量与三维打印模型评分。超低剂量组有效辐射剂量较标准剂量组降低97.1%(0.28 ± 0.05 μSv vs. 9.75 ± 2.23 μSv),两组诊断性能一致(所有评分≥2),超低剂量方案可满足三维打印模型术前评估需求。

要点

  • 超低剂量组有效辐射剂量较标准剂量组降低97.1%(0.28 ± 0.05 μSv vs. 9.75 ± 2.23 μSv)。
  • 两组三维打印模型主观评分无差异,骨折线评分标准剂量组更高(2.92±0.27 vs. 2.16 ± 0.37;p < 0.001)。
  • 两组敏感度、特异度、准确率无统计学显著差异,所有评分≥2,诊断可接受。

核心结论

  • 超低剂量方案有效降低辐射剂量97.1%,同时保持桡骨远端骨折诊断的图像质量,可满足三维打印模型术前评估需求。

研究设计与样本

  • 本研究为随机分组研究,纳入76例桡骨远端骨折患者,所有患者接受320排探测器CT扫描,随机分为标准剂量组38例(管电压120 kV、电流100 mA)和超低剂量组38例(管电压80 kV、电流10 mA)。
  • 由两名对扫描参数不知情的经验丰富骨科医生采用3分法评估三维打印模型清晰度和骨折线,评分≥2视为诊断可接受,并计算平均辐射剂量,比较两组骨折诊断性能。

关键影像表现/征象

  • 标准剂量组定量客观图像质量参数(如CNR、SNR、CT值)高于超低剂量组(p < 0.001)。
  • 两组三维打印模型主观评分无差异,但骨折线评分标准剂量组高于超低剂量组(2.92±0.27 vs. 2.16 ± 0.37;p < 0.001)。

诊断标准与关键数值

  • 采用3分法评估三维打印模型清晰度和骨折线,评分≥2视为诊断可接受;两组所有评分均≥2。
  • 标准剂量组与超低剂量组在敏感度、特异度、准确率方面无统计学显著差异(原文未报告具体数值)。

临床应用与阅片要点

  • 超低剂量CT方案(80 kV、10 mA)可有效降低辐射剂量97.1%,同时保持桡骨远端骨折诊断的图像质量,可满足三维打印模型术前评估需求。
  • 尽管超低剂量组骨折线评分低于标准剂量组(2.16 ± 0.37 vs. 2.92±0.27;p < 0.001),但所有评分≥2,诊断性能与标准剂量组一致。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To explore the effect of ultra-low-dose computed tomography (CT) on three-dimensional (3D) printing models and the diagnosis of wrist fractures.

METHOD: This study enrolled 76 patients with distal radial fractures (DRFs). All patients underwent 320-row detector CT and were divided randomly into two groups. In Group A, 38 patients were scanned with the standard-dose protocol using a tube voltage of 120 kV and current of 100 mA. In Group B, 38 patients were scanned with the ultra-low-dose protocol using a tube voltage of 80 kV and current of 10 mA. For objective image quality assessment, the noise, CT number, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured. Subjectively, two experienced orthopaedic surgeons blinded to the scan parameters evaluated the clarity of the 3D printing model and fracture line using a 3-point scale (the diagnosis was considered acceptable with scores ≥2). The mean radiation dose was calculated. The diagnostic performances for the fractures between the two groups were compared.

RESULTS: The effective radiation dose was significantly reduced by 97.1 % in Group B, compared to Group A (0.28 ± 0.05vs. 9.75 ± 2.23 μSv, respectively). Quantitative objective image quality parameters (e.g., CNR, SNR, and CT numbers) were higher in the standard-dose group (p < 0.001). However, there was no difference in subjective scoring of the 3D printing model. Although the fracture line score was higher in Group A (2.92±0.27 vs. 2.16 ± 0.37; p < 0.001), the diagnostic performance of the two groups was consistent (all scores ≥2). There were no statistically significant differences in the sensitivity, specificity or accuracy between standard-dose group and ultra-low-dose group.

CONCLUSIONS: The ultra-low-dose protocol effectively reduced the radiation dose by 97.1 %, while maintaining the image quality for diagnosis of DRFs. Therefore, this protocol can meet the needs of 3D printing models for preoperative assessments.

来源:Eur J Radiol,查看原文。版权归原作者所有。