术前CT测量的耳蜗管长度与耳蜗距离:估算人工耳蜗电极插入深度角的影像标志物
本研究纳入35例接受人工耳蜗植入的儿童,在术前CT上测量耳蜗管长度(CDL)及耳蜗基底转直径(距离A/B),并与术后X线片测得的电极插入深度角进行相关性分析。结果显示距离A与插入深度角呈负相关(r=-0.7643,p<0.0001),距离B亦呈负相关(r=-0.7118,p<0.0001),而CDL与插入深度角的相关性无统计学意义(r=-0.2333,p>0.05)。作者认为距离A可作为估算插入深度角的预测标志物。
要点
- 术前CT可重复测量CDL及距离A/B,观察者间一致性ICC分别为0.864、0.862、0.529、0.958。
- 距离A与电极插入深度角负相关(r=-0.7643,p<0.0001),距离B亦负相关(r=-0.7118,p<0.0001)。
- CDL与插入深度角相关性无统计学意义(r=-0.2333,p>0.05),作者认为距离A可用于估算插入深度角。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Preoperative estimation of the insertion depth angle of cochlear implant (CI) electrodes is essential for surgical planning. The purpose of this study was to determine the cochlear size using preoperative CT and to investigate the correlation between cochlear size and insertion depth angle in morphologically normal cochlea.
METHODS: Thirty-five children who underwent CI were included in this study. Cochlear duct length (CDL) and the diameter of the cochlear basal turn (distance A/B) on preoperative CT and the insertion depth angle of the CI electrode on postoperative radiographs were independently measured by two readers. Correlation between cochlear size and insertion depth angle was evaluated. Interobserver agreement was calculated using the intraclass correlation coefficient (ICC).
RESULTS: The mean CDL, distance A, and distance B of 70 ears were 36.20 ± 1.57 mm, 8.67 ± 0.42 mm, and 5.73 ± 0.32 mm, respectively. The mean insertion depth angle was 431.45 ± 38.42°. Interobserver agreements of CDL, distance A/B, and insertion depth angle were fair to excellent (ICC 0.864, 0.862, 0.529, and 0.958, respectively). Distance A (r = - 0.7643) and distance B (r = - 0.7118) showed a negative correlation with insertion depth angle, respectively (p < 0.0001). However, the correlation between CDL and insertion depth angle was not statistically significant (r = - 0.2333, p > 0.05).
CONCLUSIONS: The CDL and cochlear distance can be reliably obtained from preoperative CT. Distance A can be used as a predictive marker for estimating insertion depth angle during CI surgery.