临床CT中基于患者信息的器官剂量估算:在1048例临床患者中的实施与有效剂量评估
本研究在1048例患者中实施基于患者信息的器官剂量监测框架,通过将定位像匹配到58个成人、56个儿童、5个孕妇及12个ICRP参考模型组成的体模库,并结合CTDIvol和管电流调制估算器官剂量。比较基于器官剂量的有效剂量(ED)与基于剂量长度乘积(DLP)的ED,两者差异有统计学意义(p < .001),DLP-based ED对体型较大患者高估、对体型较小患者低估。
要点
- 基于患者信息的器官剂量估算框架适用于成人、儿童及孕妇的临床CT成像。
- 与基于器官剂量的ED相比,基于DLP的ED对体型较大患者可能高估有效剂量,对体型较小患者可能低估。
- 胸部扫描中DLP-based ED与器官剂量-based ED差异可达190.9%,腹盆部扫描可达234.7%(均值分别为9.0%和24.3%)。
诊断标准与关键数值
- 大多数器官的器官剂量估算最大误差:未进行偏差校正时平均为18%,进行偏差校正后平均为8%(跨所有协议平均)。
- 胸部扫描中基于DLP的ED与基于器官剂量的ED差异可达190.9%(均值9.0%),腹盆部扫描可达234.7%(均值24.3%),差异有统计学意义(p < .001)。
临床应用与阅片要点
- 在临床CT剂量监测中,可实施基于患者信息的器官剂量估算框架,该框架适用于成人、儿童和孕妇患者的成像。
- 与基于器官剂量的ED相比,基于DLP的ED可能对体型较大患者高估有效剂量、对体型较小患者低估有效剂量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this study is to comprehensively implement a patient-informed organ dose monitoring framework for clinical CT and compare the effective dose (ED) according to the patient-informed organ dose with ED according to the dose-length product (DLP) in 1048 patients. MATERIALS AND METHODS. Organ doses for a given examination are computed by matching the topogram to a computational phantom from a library of anthropomorphic phantoms and scaling the fixed tube current dose coefficients by the examination volume CT dose index (CTDIvol) and the tube-current modulation using a previously validated convolution-based technique. In this study, the library was expanded to 58 adult, 56 pediatric, five pregnant, and 12 International Commission on Radiological Protection (ICRP) reference models, and the technique was extended to include multiple protocols, a bias correction, and uncertainty estimates. The method was implemented in a clinical monitoring system to estimate organ dose and organ dose-based ED for 647 abdomen-pelvis and 401 chest examinations, which were compared with DLP-based ED using a t test. RESULTS. For the majority of the organs, the maximum errors in organ dose estimation were 18% and 8%, averaged across all protocols, without and with bias correction, respectively. For the patient examinations, DLP-based ED was significantly different from organ dose-based ED by as much as 190.9% and 234.7% for chest and abdomen-pelvis scans, respectively (mean, 9.0% and 24.3%). The differences were statistically significant (p < .001) and exhibited overestimation for larger-sized patients and underestimation for smaller-sized patients. CONCLUSION. A patient-informed organ dose estimation framework was comprehensively implemented applicable to clinical imaging of adult, pediatric, and pregnant patients. Compared with organ dose-based ED, DLP-based ED may overestimate effective dose for larger-sized patients and underestimate it for smaller-sized patients.
引用
- Fu W, Ria F, Segars WP, et al. Patient-Informed Organ Dose Estimation in Clinical CT: Implementation and Effective Dose Assessment in 1048 Clinical Patients. AJR Am J Roentgenol. 2021. PMID: 33474986. doi:10.2214/AJR.19.22482
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。