股腘动脉病变中自动二氧化碳数字减影血管造影的图像质量与安全性:一项随机单中心研究结果
这项单中心前瞻性研究纳入50例因股腘动脉外周动脉疾病接受DSA的患者,同一靶病变先后使用碘对比剂和二氧化碳作为对比剂。自动CO2-DSA联合专用后处理软件在整体图像质量、侧支可见度和狭窄/闭塞评估上与ICM-DSA相比具有fair至excellent的阅片者间一致性,且安全性可接受,但CO2-DSA的疼痛评分显著高于ICM-DSA。
要点
- CO2-DSA与ICM-DSA的阅片者间一致性在整体图像质量上为fair至excellent(ICC 0.399-0.748)。
- CO2-DSA联合自动注射系统与专用后处理软件的安全性可接受,未观察到其他不良事件。
关键影像表现/征象
- CO2-DSA与ICM-DSA在整体图像质量上的阅片者间一致性为fair至excellent(ICC 0.399-0.748)。
- CO2-DSA与ICM-DSA在侧支可见度上的阅片者间一致性为fair至excellent(ICC 0.513-0.691)。
- CO2-DSA与ICM-DSA在狭窄/闭塞评估上的阅片者间一致性为poor至excellent(ICC -0.065-0.762)。
- 各阅片者之间无显著差异。
诊断标准与关键数值
- 由四名阅片者独立基于整体图像质量、侧支可见度和狭窄/闭塞评估对血管造影图像进行评价。
- 阅片者间一致性采用组内相关系数(ICC)评估,阅片者间差异采用Friedmann检验评估。
临床应用与阅片要点
- CO2-DSA可作为股腘动脉外周动脉疾病DSA检查中与ICM-DSA相比可比的诊断检查。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the image quality and the safety of automated carbon dioxide (CO2) digital subtraction angiography (DSA).
MATERIALS AND METHODS: Fifty patients receiving DSA for femoropopliteal peripheral arterial disease (PAD) were enrolled in this single-center prospective study. All patients received iodinated contrast media (ICM) and CO2 as a contrast agent in the same target lesion. As a primary endpoint, four raters independently evaluated the angiography images based on overall image quality, visibility of collaterals, and assessment of stenoses/occlusions. Inter-rater agreement was assessed using the intraclass correlation coefficient (ICC) and differences between the raters were evaluated using Friedmann's test. Secondary endpoints were procedure safety and patient pain assessment.
RESULTS: Inter-rater agreement between CO2-DSA and ICM-DSA images was fair to excellent for overall image quality (ICC: 0.399-0.748), fair to excellent for the visibility of collaterals (ICC: 0.513-0.691), and poor to excellent for the assessment of stenoses/occlusions (ICC: -0.065-0.762). There were no significant differences between the raters. Two patients had a hematoma, one reported pain related to puncture, one became nauseous, and one vomited. No other adverse events were observed. Reported pain scores were significantly higher for CO2-DSA vs. ICM-DSA (1.25 vs. 0625; p < 0.028).
CONCLUSION: CO2-DSA using automated injection system in combination with proprietary post-processing software is safe and comparable diagnostic test compared to ICM-DSA.
引用
- Bürckenmeyer F, Schmidt A, Diamantis I, et al. Image quality and safety of automated carbon dioxide digital subtraction angiography in femoropopliteal lesions: Results from a randomized single-center study. Eur J Radiol. 2020. PMID: 33388532. doi:10.1016/j.ejrad.2020.109476
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。