自由手三维超声测量人跟腱硬度的可靠性

Reliability of Human Achilles Tendon Stiffness Measures Using Freehand 3-D Ultrasound.
PubMed 文献 · Ultrasound Med Biol · 2021-01-21

研究纳入10名健康受试者,在同一天间隔1小时两次扫描自由跟腱(距跟骨2-6厘米),于静息及最大等长跖屈力的20%、40%、60%收缩状态下测量硬度。自由手三维超声测量的重测信度优秀(ICC=0.994,95%CI:0.978-0.999),测试与重测均值无显著差异(p=0.72),最小可检测变化为52.14 N/mm,Bland-Altman图未显示系统偏倚(95%CI:-22.18至15.88)。

要点

  • 自由手三维超声测量自由跟腱硬度的重测信度优秀,ICC=0.994(95%CI:0.978-0.999)。
  • 测试均值361.83 N/mm(170.77)与重测均值364.98 N/mm(168.57)无显著差异(p=0.72)。
  • 最小可检测变化为52.14 N/mm,Bland-Altman图未显示系统偏倚(95%CI:-22.18至15.88)。

核心结论

  • 自由手三维超声可提供可靠且精确的跟腱硬度测量,能够检测自由跟腱硬度随负荷或腱病发生的微小变化。

研究设计与样本

  • 方法学信度研究,纳入健康受试者10名,同一天间隔1小时进行两次自由手三维超声扫描自由跟腱(距跟骨2-6厘米)。
  • 于静息及最大等长跖屈力的20%、40%、60%收缩状态下测量,以力-伸长曲线斜率作为个体硬度(N/mm)。
  • 相对信度采用组内相关系数(ICC)评估,绝对信度采用测量标准误(SEM)和最小可检测变化评估,并通过比较测试与重测分布及Bland-Altman图探索系统偏倚。

诊断标准与关键数值

  • 自由跟腱硬度重测信度ICC=0.994(95%CI:0.978-0.999)。
  • 测试均值硬度361.83 N/mm(170.77)与重测均值364.98 N/mm(168.57)无显著差异(p=0.72)。
  • 最小可检测变化为52.14 N/mm;Bland-Altman图未显示系统偏倚(95%CI:-22.18至15.88)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Achilles tendon (AT) stiffness is an important property of both human locomotor performance and injury mechanics. Freehand 3-D ultrasound (3-DUS) is a promising method for measuring stiffness of the Achilles tendon, particularly the free AT (2-6 cm proximal to calcaneus), which is commonly injured. The aim of this study was to investigate the test-retest reliability of freehand 3-DUS in measuring free AT stiffness in humans. The free Achilles tendon length of healthy participants (n = 10) was scanned on the same day on two consecutive occasions (1 h apart) during rest and isometric plantar flexion contractions at 20%, 40% and 60% of maximum force. The slope of the force-elongation curve over these force levels represented individual stiffness (N/mm). Relative reliability was assessed using the intra-class correlation coefficient (ICC), and absolute reliability was estimated with the standard error of measurement (SEM) and smallest detectable change. Systematic bias in stiffness measures was explored by comparing test and retest distributions and Bland-Altman plots. The test-retest reliability of free AT stiffness measured using freehand 3-DUS was excellent [ICC = 0.994, 95% confidence interval [CI]: 0.978-0.999)]. The mean stiffness values at test (361.83 N/mm [170.77]) and retest (364.98 N/mm [168.57]) did not significantly differ (p = 0.72), and the smallest detectable change was 52.14 N/mm. The Bland-Altman plot indicated the absence of systematic bias (95% CI: -22.18 to 15.88). Freehand 3-DUS provides reliable and precise measures of tendon stiffness and can be used to detect small changes in free AT stiffness in response to load or tendon pathology.