确定射频消融治疗良性甲状腺结节实现最佳体积缩减的能量阈值

Determining an energy threshold for optimal volume reduction of benign thyroid nodules treated by radiofrequency ablation.
PubMed 文献 · Eur Radiol · 2021-01-06

这项回顾性研究纳入三家机构按同一方案治疗的115例良性甲状腺结节患者,分析超声与技术参数对1年体积缩减率(VRR)的影响。基线体积、总能量和单位体积能量与VRR>50%独立相关,概率回归显示单位体积能量756 J/ml和2670 J/ml时VRR>50%的概率分别为50%和99%。作者认为结合基线结节体积并给予足够的单位体积能量可优化技术成功与临床成功。

要点

  • 基线结节体积、总能量和单位体积能量是VRR>50%的独立影响因素。
  • 单位体积能量756 J/ml时VRR>50%的概率为50%,2670 J/ml时概率为99%。
  • 作者提出总能量可按结节体积×2670 J估算。

核心结论

  • 考虑基线结节体积并给予足够的单位体积能量,可优化射频消融治疗良性甲状腺结节的技术成功与临床成功。

研究设计与样本

  • 回顾性研究,三家机构按同一方案治疗,共纳入115例良性甲状腺结节患者。
  • 技术参数为功率和能量,以1年体积缩减率(VRR)作为疗效参考标准。
  • 采用单因素和多因素分析评估参数与VRR>50%的关联,并用概率回归估计实现有效反应的概率。

关键影像表现/征象

  • 中位功率为50 W,中位总释放能量为27,531 J。
  • 1年随访时中位VRR为64.0%,87个(75.7%)结节VRR>50%。

诊断标准与关键数值

  • 基线体积、总能量、单位体积能量与VRR>50%独立相关,P值分别为0.001、0.0178和<0.001。
  • 概率回归分析显示,单位体积能量756 J/ml时VRR>50%的概率为50%,2670 J/ml时概率为99%。

临床应用与阅片要点

  • 作者提出规划治疗时总能量可按公式「结节体积×2670 J」计算。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Radiofrequency ablation (RFA) is effective in reducing the volume of benign thyroid nodules. However, what parameters can influence the response to RFA is still unclear. The present study aimed to (1) investigate which ultrasound and technical parameters are potential determinants of the volumetric reduction; (2) develop a dose-response model, and (3) analyze the effects of RFA on ultrasound features.

METHODS: In this retrospective study, three institutions treated patients with benign thyroid nodules according to the same protocol. The technical parameters were power and energy. The 1-year volume reduction ratio (VRR) was the reference standard of the response. The correlations of different parameters with VRR were analyzed and the association between several parameters and a VRR above 50% studied by uni- and multivariate analyses. The probit regression estimated the probability to achieve an effective response.

RESULTS: One hundred fifteen patients were enrolled. The median power was 50 W and median total delivered energy 27,531 J. At 1-year follow-up, the median VRR was 64.0% and 87 (75.7%) nodules showed a VRR above 50%. Among all parameters, only baseline volume, total energy, and energy per volume were independently associated to a VRR > 50% (p = 0.001, p = 0.0178, p < 0.001 respectively). The probit regression analysis demonstrated that delivering 756 J/ml and 2670 J/ml gave a probability of VRR > 50% in 50% and 99% of patients, respectively.

CONCLUSIONS: Considering the baseline nodular volume and delivering the adequate energy per volume allow optimizing technical and clinical success.

KEY POINTS: • The effectiveness of radiofrequency ablation in treating benign thyroid nodules is negatively correlated to the volume of the nodule and positively correlated to the energy delivered per volume. • When planning the treatment, the total energy to deliver can be calculated by using a simple formula: nodular volume × 2670 J.

来源:Eur Radiol,查看原文。版权归原作者所有。