文献速递外周血管介入技术与安全CTDSA

血液透析导管在计算机断层扫描血管造影机械压力注射碘对比剂时的完整性

Hemodialysis catheter integrity during mechanical power injection of iodinated contrast medium for computed tomography angiography.
PubMed 文献 · Abdom Radiol (NY) · 2021-01-02

本研究为体外研究,评估4种血液透析导管在不同对比剂类型、温度、导管直径、长度及单/双腔同时注射条件下,经机械压力注射器给药的結構完整性。共864次注射,无导管破裂;双腔同时注射峰值压力26.3 psi,单腔注射51.7 psi,双腔注射及低黏度对比剂峰值压力显著更低。结论认为体外条件下血液透析导管可安全用于机械压力注射碘对比剂。

要点

  • 体外864次注射中未见导管膨出或破裂。
  • 双腔同时注射峰值压力26.3 psi,单腔注射51.7 psi。
  • 双腔注射及低黏度对比剂峰值压力显著更低(p < 0.001)。

核心结论

  • 体外操作条件下,血液透析导管可经机械压力注射器安全给予碘对比剂。
  • 双腔注射流速达20 mL/s、单腔注射流速达10 mL/s时,最大峰值压力低于制造商30 psi限值,均高于临床CT血管造影常用最大流速8 mL/s。

研究设计与样本

  • 体外研究,评估4种血液透析导管。
  • 测试变量包括对比剂类型(碘帕醇300和370 mgI/mL)、温度(25和37 °C)、导管直径(14 Fr至16 Fr,均为双腔)、导管长度(19-32 cm)及单腔或双腔同时注射。
  • 以5至20 mL/s流速(5 mL/s递增)记录峰值平台压力(psi),共完成864次独立注射。

诊断标准与关键数值

  • 864次注射中无导管失败(膨出/破裂)发生。
  • 单腔注射最大压力51.7 psi,双腔注射26.3 psi。
  • 双腔同时注射较单腔注射峰值压力显著更低(p < 0.001)。
  • 低黏度对比剂较高黏度对比剂峰值压力显著更低(p < 0.001)。
  • 较大与较小直径管腔间无显著差异(p = 0.221),单腔注射动脉与静脉途径间无显著差异(p = 0.834)。

临床应用与阅片要点

  • 体外条件下,双腔注射流速至20 mL/s、单腔注射流速至10 mL/s时峰值压力低于制造商30 psi限值。
  • 上述流速均高于临床CT血管造影常用最大流速8 mL/s。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: CT angiography (CTA) requires vascular access with flow rates of 5-7 mL/s. Hemodialysis (HD) is performed at 6-10 mL/s. The purpose of our study is to evaluate the structural integrity of HD catheters in the administration of contrast media via a mechanical power injector under varying conditions.

METHODS: Four HD catheters were evaluated in an in vitro study. Tested were contrast media type (iopamidol 300 and 370 mgI/mL), temperature (25 and 37 °C), catheter diameter (14 Fr to 16 Fr all with double-lumen capacity), catheter length (19-32 cm), and simultaneous double-lumen or single-lumen injection within each of the catheters. Peak plateau pressures (psi) were recorded with flow rates from 5 to 20 mL/s in 5 mL/s increments. In total, 864 unique injections were performed.

RESULTS: No catheter failure (bulging/rupture) was observed in 864 injections. Maximum pressure for single-lumen injection was 51.7 psi (double-lumen: 26.3 psi). Peak pressures were significantly lower in simultaneous double-lumen vs. single-lumen injections (p < 0.001) and low vs. high viscosity contrast media (p < 0.001). Neither larger vs. smaller diameter lumens (p = 0.221) nor single-lumen injection in arterial vs. venous (p = 0.834) were significantly different.

CONCLUSION: HD catheters can be used to safely administer iodinated contrast media via mechanical power injection in in vitro operating conditions. Maximum peak pressure is below the manufacturer's 30 psi limit at flow rates up to 20 mL/s in double-lumen injections and up to 10 mL/s in single-lumen injections, which is higher than the usual maximum of 8 mL/s for CT angiography in clinical settings.