文献速递腹部技术与安全CT

四种双能CT平台腹部成像性能:基于体模数据的任务型图像质量评估

Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data.
PubMed 文献 · Eur Radiol · 2021-01-15

研究用两个体模在快kV切换、分裂滤片、双源和双层四种双能CT平台上扫描,以10 mGy的CTDIvol和腹部盆腔经典参数采集,评估虚拟单能图像40–140 keV的噪声功率谱、任务型传递函数和检出指数。结果显示双层CT和双源CT噪声最低,双层CT和双源CT对两种模拟对比增强病灶的检出性最高,除快kV切换平台在70 keV外其余平台均在40 keV检出性最高。

要点

  • 双层CT和双源CT的噪声幅度最低,快kV切换和分裂滤片平台噪声最高。
  • 双层CT和双源CT对两种模拟对比增强病灶的检出指数最高。
  • 除快kV切换平台在70 keV检出性最高外,其余平台均在40 keV检出性最高。

关键影像表现/征象

  • 所有双能CT平台从40到70 keV噪声幅度下降,快kV切换CT和分裂滤片CT噪声幅度值最高,双源CT和双层CT最低。
  • 所有双能CT平台随能级增加平均噪声功率谱空间频率向低频移动,图像纹理变平滑。
  • 任务型传递函数值随keV增加而下降,空间分辨率变差。

诊断标准与关键数值

  • 检出指数d'用于模拟两种对比增强病灶的检测任务,作为keV的函数。
  • 双层CT、双源CT和分裂滤片CT在40 keV获得较高检出指数,快kV切换CT在70 keV获得较高检出指数。
  • 两种模拟病灶的检出性在双层CT和双源CT最高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the spectral performance of dual-energy CT (DECT) platforms using task-based image quality assessment based on phantom data.

MATERIALS AND METHODS: Two CT phantoms were scanned on four DECT platforms: fast kV-switching CT (KVSCT), split filter CT (SFCT), dual-source CT (DSCT), and dual-layer CT (DLCT). Acquisitions on each phantom were performed using classical parameters of abdomen-pelvic examination and a CTDIvol at 10 mGy. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated from 40 to 140 keV of virtual monoenergetic images. A detectability index (d') was computed to model the detection task of two contrast-enhanced lesions as function of keV.

RESULTS: The noise magnitude decreased from 40 to 70 keV for all DECT platforms, and the highest noise magnitude values were found for KVSCT and SFCT and the lowest for DSCT and DLCT. The average NPS spatial frequency shifted towards lower frequencies as the energy level increased for all DECT platforms, smoothing the image texture. TTF values decreased with the increase of keV deteriorating the spatial resolution. For both simulated lesions, higher detectability (d' value) was obtained at 40 keV for DLCT, DSCT, and SFCT but at 70 keV for KVSCT. The detectability of both simulated lesions was highest for DLCT and DSCT.

CONCLUSION: Highest detectability was found for DLCT for the lowest energy levels. The task-based image quality assessment used for the first time for DECT acquisitions showed the benefit of using low keV for the detection of contrast-enhanced lesions.

KEY POINTS: • Detectability of both simulated contrast-enhanced lesions was higher for dual-layer CT for the lowest energy levels. • The image noise increased and the image texture changed for the lowest energy levels. • The detectability of both simulated contrast-enhanced lesions was highest at 40 keV for all dual-energy CT platforms except for fast kV-switching platform.