深度学习训练算法可维持半剂量对比增强肝脏CT图像质量:与混合迭代重建的比较:深度学习降噪技术在低剂量中的应用研究

Deep learning trained algorithm maintains the quality of half-dose contrast-enhanced liver computed tomography images: Comparison with hybrid iterative reconstruction: Study for the application of deep learning noise reduction technology in low dose.
PubMed 文献 · Eur J Radiol · 2021-01-05

本研究纳入207名成人,比较半剂量对比增强肝脏CT采用深度学习训练算法(DELTA)与标准剂量CT采用混合迭代重建(HIR)的图像及诊断质量。结果显示LDCTDL噪声显著低于SDCTHIR和LDCTHIR,SNR和CNR显著更高,整体图像质量、解剖结构显示及病灶显示与SDCTHIR无显著差异。结论为LDCT联合DELTA较SDCT联合HIR剂量降低约49%而维持图像质量。

要点

  • 半剂量肝脏CT采用DELTA重建,有效剂量2.87 mSv较标准剂量5.64 mSv降低约49%。
  • LDCTDL噪声显著低于SDCTHIR和LDCTHIR,SNR和CNR显著高于两者。
  • LDCTDL与SDCTHIR在整体图像质量、解剖结构显示及病灶显示上无显著差异。

核心结论

  • LDCT联合DELTA较SDCT联合HIR剂量降低约49%,同时维持对比增强肝脏CT的图像质量。

研究设计与样本

  • 本研究纳入207名成人,分为标准剂量CT(SDCT)组和低剂量CT(LDCT)组。
  • SDCT采用混合迭代重建(HIR)生成SDCTHIR,LDCT分别采用HIR(LDCTHIR)和DELTA(LDCTDL)重建。
  • 比较三组图像的噪声、CT值(HU)、信噪比(SNR)和对比噪声比(CNR),并由两名放射科医师评估噪声、伪影、整体图像质量、关键解剖结构显示及病灶检测、表征和显示。

诊断标准与关键数值

  • 平均有效剂量:SDCT为5.64 ± 1.96 mSv,LDCT为2.87 ± 0.87 mSv。
  • 病灶检测(基于患者)的敏感度和特异度:SDCTHIR分别为81.9%和89.1%,LDCTDL分别为83.7%和86.3%。
  • 病灶表征(基于患者)的敏感度和特异度:SDCTHIR分别为75.4%和82.6%,LDCTDL分别为73.5%和82.4%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: This study compares the image and diagnostic qualities of a DEep Learning Trained Algorithm (DELTA) for half-dose contrast-enhanced liver computed tomography (CT) with those of a commercial hybrid iterative reconstruction (HIR) method used for standard-dose CT (SDCT).

METHODS: This study enrolled 207 adults, and they were divided into two groups: SDCT and low-dose CT (LDCT). SDCT was reconstructed using the HIR method (SDCTHIR), and LDCT was reconstructed using both the HIR method (LDCTHIR) and DELTA (LDCTDL). Noise, Hounsfield unit (HU) values, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were compared between three image series. Two radiologists assessed the noise, artefacts, overall image quality, visualisation of critical anatomical structures and lesion detection, characterisation and visualisation.

RESULTS: The mean effective doses were 5.64 ± 1.96 mSv for SDCT and 2.87 ± 0.87 mSv for LDCT. The noise of LDCTDL was significantly lower than that of SDCTHIR and LDCTHIR. The SNR and CNR of LDCTDL were significantly higher than those of the other two groups. The overall image quality, visualisation of anatomical structures and lesion visualisation between LDCTDL and SDCTHIR were not significantly different. For lesion detection, the sensitivities and specificities of SDCTHIR vs. LDCTDL were 81.9 % vs. 83.7 % and 89.1 % vs. 86.3 %, respectively, on a per-patient basis. SDCTHIR showed 75.4 % sensitivity and 82.6 % specificity for lesion characterisation on a per-patient basis, whereas LDCTDL showed 73.5 % sensitivity and 82.4 % specificity.

CONCLUSIONS: LDCT with DELTA had approximately 49 % dose reduction compared with SDCT with HIR while maintaining image quality on contrast-enhanced liver CT.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Zeng L, Xu X, Zeng W, et al. Deep learning trained algorithm maintains the quality of half-dose contrast-enhanced liver computed tomography images: Comparison with hybrid iterative reconstruction: Study for the application of deep learning noise reduction technology in low dose. Eur J Radiol. 2021. PMID: 33418383. doi:10.1016/j.ejrad.2020.109487