北美放射学会/定量成像生物标志物联盟在弹性和黏弹性体模中剪切波速度偏倚的量化研究

Radiological Society of North America/Quantitative Imaging Biomarker Alliance Shear Wave Speed Bias Quantification in Elastic and Viscoelastic Phantoms.
PubMed 文献 · J Ultrasound Med · 2021-01-07

该研究用弹性和黏弹性体模量化不同商用超声剪切弹性系统与磁共振弹性成像(MRE)系统间剪切波速度(SWS)测量的偏倚。弹性体模用5个超声平台、黏弹性体模用12个超声平台测量,不同系统间SWS平均差异的95%置信区间在弹性体模为±0.145 m/s(±9.6%)、黏弹性体模为±0.340 m/s(±15.3%);MRE在黏弹性体模中以140 Hz激励时与超声SWS最匹配,但临床所用60 Hz振动源存在-0.27±0.027 m/s(-12.2%±1.2%)的偏倚。

要点

  • 不同厂商超声系统间SWS重建在弹性体模中较黏弹性体模更一致,所有情况平均差异偏倚<±10%。
  • 黏弹性体模中MRE以140 Hz激励时与超声SWS最匹配,而临床60 Hz振动源存在显著负偏倚。
  • 焦距和评估者对SWS变异的影响小于系统和站点。

诊断标准与关键数值

  • 不同系统间SWS平均差异的95%置信区间:弹性体模为±0.145 m/s(±9.6%),黏弹性体模为±0.340 m/s(±15.3%)。
  • MRE在黏弹性体模中以140 Hz激励时与超声SWS最匹配;使用临床实施的60 Hz振动源时偏倚为-0.27±0.027 m/s(-12.2%±1.2%)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To quantify the bias of shear wave speed (SWS) measurements between different commercial ultrasonic shear elasticity systems and a magnetic resonance elastography (MRE) system in elastic and viscoelastic phantoms.

METHODS: Two elastic phantoms, representing healthy through fibrotic liver, were measured with 5 different ultrasound platforms, and 3 viscoelastic phantoms, representing healthy through fibrotic liver tissue, were measured with 12 different ultrasound platforms. Measurements were performed with different systems at different sites, at 3 focal depths, and with different appraisers. The SWS bias across the systems was quantified as a function of the system, site, focal depth, and appraiser. A single MRE research system was also used to characterize these phantoms using discrete frequencies from 60 to 500 Hz.

RESULTS: The SWS from different systems had mean difference 95% confidence intervals of ±0.145 m/s (±9.6%) across both elastic phantoms and ± 0.340 m/s (±15.3%) across the viscoelastic phantoms. The focal depth and appraiser were less significant sources of SWS variability than the system and site. Magnetic resonance elastography best matched the ultrasonic SWS in the viscoelastic phantoms using a 140 Hz source but had a - 0.27 ± 0.027-m/s (-12.2% ± 1.2%) bias when using the clinically implemented 60-Hz vibration source.

CONCLUSIONS: Shear wave speed reconstruction across different manufacturer systems is more consistent in elastic than viscoelastic phantoms, with a mean difference bias of < ±10% in all cases. Magnetic resonance elastographic measurements in the elastic and viscoelastic phantoms best match the ultrasound systems with a 140-Hz excitation but have a significant negative bias operating at 60 Hz. This study establishes a foundation for meaningful comparison of SWS measurements made with different platforms.