质子密度脂肪分数作为定量影像生物标志物的线性与偏倚:多中心、多平台、多厂商体模研究

Linearity and Bias of Proton Density Fat Fraction as a Quantitative Imaging Biomarker: A Multicenter, Multiplatform, Multivendor Phantom Study.
PubMed 文献 · Radiology · 2021-01-19

本前瞻性研究将含12个已知质子密度脂肪分数(PDFF)值小瓶的商用体模运送至美国9个中心,在3个厂商的27台1.5T和3.0T系统上完成160次独立MRI检查,比较两种三维化学位移编码MRI协议估算的PDFF与已知值的线性与偏倚。结果显示在0%-47.5% PDFF范围内,汇总回归斜率为0.97(95%CI:0.96,0.98),截距为-0.27%(95%CI:-0.50%,-0.05%),表明基于厂商产品采集与重建软件的PDFF估算在不同厂商、中心和场强间准确。

要点

  • 在0%-47.5% PDFF范围内,汇总回归斜率0.97(95%CI:0.96,0.98)、截距-0.27%(95%CI:-0.50%,-0.05%)。
  • 各厂商斜率范围:1.5T厂商协议0.86-1.02、标准化协议0.97-1.0;3.0T厂商协议0.91-1.01、标准化协议0.87-1.01。
  • 西门子重建需更改配置以纠正体模中的水脂交换。

核心结论

  • 基于三维化学位移编码MRI的PDFF估算在商用体模中跨厂商、影像中心和场强均准确,使用厂商产品采集与重建软件。

研究设计与样本

  • 前瞻性研究,商用体模含12个已知PDFF值小瓶,运送至美国9个中心,在3个厂商的27台1.5T和3.0T系统上完成160次独立MRI检查。
  • 采用两种三维化学位移编码MRI协议(厂商协议和标准化协议),均具有最小T1偏倚,使用各厂商的混杂因素校正复数或混合幅值-复数重建算法生成PDFF图。
  • 以体模已知PDFF值为参考,采用线性回归混合效应模型比较MRI PDFF值,计算回归斜率(比例偏倚)和截距(恒定偏倚)的95%CI,并与无效假设(斜率=1,截距=0)比较。

诊断标准与关键数值

  • 在0%-47.5% PDFF范围内,汇总回归斜率为0.97(95%CI:0.96,0.98),汇总截距为-0.27%(95%CI:-0.50%,-0.05%)。
  • 1.5T厂商协议斜率范围0.86-1.02,标准化协议0.97-1.0;3.0T厂商协议0.91-1.01,标准化协议0.87-1.01。
  • 1.5T厂商协议截距范围-0.65%至0.18%,标准化协议-0.69%至-0.17%;3.0T厂商协议-0.48%至0.10%,标准化协议-0.78%至-0.21%(绝对PDFF百分比)。

临床应用与阅片要点

  • 使用厂商产品采集与重建软件时,三维化学位移编码MRI估算的PDFF在体模中跨厂商、中心和场强准确。
  • 西门子重建需更改配置以纠正体模中的水脂交换。
  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background Proton density fat fraction (PDFF) estimated by using chemical shift-encoded (CSE) MRI is an accepted imaging biomarker of hepatic steatosis. This work aims to promote standardized use of CSE MRI to estimate PDFF. Purpose To assess the accuracy of CSE MRI methods for estimating PDFF by determining the linearity and range of bias observed in a phantom. Materials and Methods In this prospective study, a commercial phantom with 12 vials of known PDFF values were shipped across nine U.S. centers. The phantom underwent 160 independent MRI examinations on 27 1.5-T and 3.0-T systems from three vendors. Two three-dimensional CSE MRI protocols with minimal T1 bias were included: vendor and standardized. Each vendor's confounder-corrected complex or hybrid magnitude-complex based reconstruction algorithm was used to generate PDFF maps in both protocols. The Siemens reconstruction required a configuration change to correct for water-fat swaps in the phantom. The MRI PDFF values were compared with the known PDFF values by using linear regression with mixed-effects modeling. The 95% CIs were calculated for the regression slope (ie, proportional bias) and intercept (ie, constant bias) and compared with the null hypothesis (slope = 1, intercept = 0). Results Pooled regression slope for estimated PDFF values versus phantom-derived reference PDFF values was 0.97 (95% CI: 0.96, 0.98) in the biologically relevant 0%-47.5% PDFF range. The corresponding pooled intercept was -0.27% (95% CI: -0.50%, -0.05%). Across vendors, slope ranges were 0.86-1.02 (vendor protocols) and 0.97-1.0 (standardized protocol) at 1.5 T and 0.91-1.01 (vendor protocols) and 0.87-1.01 (standardized protocol) at 3.0 T. The intercept ranges (absolute PDFF percentage) were -0.65% to 0.18% (vendor protocols) and -0.69% to -0.17% (standardized protocol) at 1.5 T and -0.48% to 0.10% (vendor protocols) and -0.78% to -0.21% (standardized protocol) at 3.0 T. Conclusion Proton density fat fraction estimation derived from three-dimensional chemical shift-encoded MRI in a commercial phantom was accurate across vendors, imaging centers, and field strengths, with use of the vendors' product acquisition and reconstruction software. © RSNA, 2021 See also the editorial by Dyke in this issue.