定量双能CT的重复扫描间与扫描仪间变异:三种不同扫描仪类型的评估

Inter-scan and inter-scanner variation of quantitative dual-energy CT: evaluation with three different scanner types.
PubMed 文献 · Eur Radiol · 2021-01-14

回顾性纳入131例在同一扫描仪上接受两次腹部双能CT检查的患者(双源DECT 46例、快速kV切换DECT 45例、双层探测器DECT 40例)。结果显示碘浓度的重复扫描间变异显著高于65keV虚拟单能图像衰减值,且扫描仪间纵向变异差异取决于扫描仪组合与评估器官。

要点

  • 碘浓度的重复扫描间变异显著高于65keV衰减值(p<0.05),阅片时不宜将碘浓度视为与CT值同等稳定的定量指标。
  • 肝脏碘浓度重复扫描间变异:快速kV切换DECT显著低于双源DECT(p=0.005)和双层探测器DECT(p=0.01)。
  • 将肝脏碘浓度归一化至门静脉、脾脏碘浓度归一化至主动脉并未显著降低重复扫描间变异(p=0.97和p=0.50)。

核心结论

  • 不同DECT扫描仪的碘测量存在重复扫描间变异,且该变异高于衰减值的变异。
  • 纵向变异与总体碘浓度的扫描仪间差异取决于所比较的扫描仪组合及评估的器官,临床与科研DECT应用中应予以认识。

研究设计与样本

  • 回顾性纳入131例在同一扫描仪上接受两次腹部DECT检查的患者:双源DECT 46例、快速kV切换DECT 45例、双层探测器DECT 40例。
  • 在肝脏、脾脏、肾脏、主动脉、门静脉和下腔静脉放置感兴趣区,测量碘浓度与65keV虚拟单能图像衰减值(HU65keV)。
  • 计算每种扫描仪的总体碘浓度与HU65keV、其重复扫描间差异及比例变异,并在扫描仪类型间比较。

关键影像表现/征象

  • 三种扫描仪队列在体重、体径、年龄、性别及对比剂注射参数及其重复扫描间差异方面相似(p范围0.23–0.99)。
  • HU65keV与碘浓度的绝对重复扫描间差异在扫描仪间相当(p范围0.08–1.0)。

诊断标准与关键数值

  • 总体重复扫描间变异碘浓度显著高于HU65keV(p<0.05)。
  • 肝脏碘浓度重复扫描间变异:快速kV切换DECT显著低于双源DECT(p=0.005)和双层探测器DECT(p=0.01)。
  • 脾脏碘浓度重复扫描间变异:快速kV切换DECT仅与双源DECT相比差异显著(p=0.015)。
  • 肝脏碘浓度归一化至门静脉、脾脏碘浓度归一化至主动脉未显著降低重复扫描间变异(p=0.97和p=0.50)。

临床应用与阅片要点

  • 临床与科研DECT应用中应认识碘测量的重复扫描间变异高于衰减值变异,纵向比较碘浓度时需考虑扫描仪类型与器官差异。
  • 三种扫描仪在总体碘浓度与两个时间点衰减值上的差异,对多数评估器官与血管而言处于患者内平均重复扫描间差异范围内。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate inter-scan and inter-scanner variation of iodine concentration (IC) and attenuation in virtual monoenergetic images at 65 keV (HU65keV) in patients with repeated abdominal examinations on dual-source (dsDECT), rapid kV switching (rsDECT), and dual-layer detector DECT (dlDECT).

METHODS: We retrospectively included 131 patients who underwent two abdominal DECT examinations on the same scanner (dsDECT: n = 46, rsDECT: n = 45, dlDECT: n = 40). IC and HU65keV were measured by placing regions of interest in the liver, spleen, kidneys, aorta, portal vein, and inferior vena cava. Overall IC and HU65keV for each scanner, their inter-scan differences and proportional variation were calculated and compared between scanner types.

RESULTS: The three scanner-specific cohorts showed similar weight, body diameter, age, sex, and contrast media injection parameters as well as inter-scan differences hereof (p range: 0.23-0.99). Absolute inter-scan differences of HU65keV and IC were comparable between scanners (p range: 0.08-1.0). Overall inter-scan variation was significantly higher in IC than HU65keV (p < 0.05). For the liver, rsDECT showed significantly lower inter-scan variation of IC compared to dsDECT/dlDECT (p = 0.005/0.01), while for the spleen, this difference was only significant compared to dsDECT (p = 0.015). Normalizing IC of the liver to the portal vein and of the spleen to the aorta did not significantly reduce inter-scan variation (p = 0.97 and 0.50).

CONCLUSIONS: Iodine measurements across different DECT scanners show inter-scan variation which is higher compared to variation of attenuation values. Inter-scanner differences in longitudinal variation and overall iodine concentration depend on the scanner pairs and organs assessed and should be acknowledged in clinical and scientific DECT applications.

KEY POINTS: • All scanner types showed comparable inter-scan variation of attenuation, while for iodine, the rapid kV switching DECT showed lower variability in the liver and spleen. • Iodine concentration showed higher inter-scan variation than attenuation measurements; normalization to vessels did not significantly improve inter-scan reproducibility of iodine concentration in parenchymal organs. • Differences between the three scanner types regarding overall iodine concentration and attenuation obtained from both timepoints were within the range of average intra-patient, inter-scan differences for most assessed organs and vessels.

来源:Eur Radiol,查看原文。版权归原作者所有。