心脏CT金属伪影减少算法评估心脏植入式电子装置导线穿孔

Assessment of cardiac implantable electric device lead perforation using a metal artifact reduction algorithm in cardiac computed tomography.
PubMed 文献 · Eur J Radiol · 2021-01-09

回顾性纳入47例心脏植入式电子装置患者,每例扫描分别用常规算法和金属伪影减少(MAR)算法重建,比较右心室导线尖端定位。MAR算法伪影指数显著低于常规算法(96.7±40.1 HU vs 284.6±134.1 HU,P<0.001),不确定诊断率由27.7%降至0%(P<0.001),全部病例可判定穿孔(6例,12.8%)或非穿孔(41例,87.2%)。

要点

  • MAR算法将伪影指数从284.6±134.1 HU降至96.7±40.1 HU(P<0.001)。
  • 常规算法下27.7%病例无法定性,MAR算法下全部病例可判定穿孔与否。
  • MAR算法判定穿孔6例(12.8%)、非穿孔41例(87.2%)。

核心结论

  • MAR算法有效减少金属伪影,使全部病例均可诊断有无导线穿孔,而常规算法下27.7%的病例难以明确诊断。

研究设计与样本

  • 连续纳入47例(男26例,年龄70.3±15.4岁)心脏植入式电子装置患者行心脏CT,每例扫描分别用常规算法和MAR算法重建,计算伪影指数并视觉评估右心室导线尖端位置(非穿孔、穿孔、不确定)。

关键影像表现/征象

  • 常规算法下13例(27.7%)右心室导线尖端位置判为不确定,MAR算法下不确定率为0%(P<0.001)。
  • MAR算法将常规算法判为不确定的13例重新判为穿孔2例、非穿孔11例。
  • MAR算法下全部病例判为穿孔6例(12.8%)或非穿孔41例(87.2%)。

诊断标准与关键数值

  • 伪影指数均值MAR算法96.7±40.1 HU vs 常规算法284.6±134.1 HU,MAR算法更低(P<0.001)。
  • 不确定诊断率MAR算法0% vs 常规算法27.7%(P<0.001)。
  • 右心室导线尖端位置判读分为非穿孔、穿孔、不确定三类。

临床应用与阅片要点

  • 常规算法重建出现导线尖端伪影致位置不确定时,可加做MAR算法重建辅助判定有无穿孔。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: CT is considered the non-invasive gold standard for evaluating cardiac implantable electronic devices (CIEDs) lead perforation, but metal artifacts caused by the lead tip affect the image quality and make a definitive diagnosis challenging. We compared the performances of the metal artifact reduction (MAR) algorithm and the conventional algorithm for identification of the right ventricular (RV) lead tip position in cardiac CT studies of patients with CIEDs.

METHOD: Forty-seven consecutive patients (26 men; age 70.3 ± 15.4 years) with CIEDs underwent cardiac CT. Using the conventional and MAR algorithm, two image reconstructions were performed for each scan. We calculated the artifact index (AI) to assess the quantitative capability of the MAR algorithm for artifact reduction and visually assessed the RV lead tip position on both images as follows: non-perforation, perforation, and equivocal.

RESULTS: The mean AIs were significantly lower with the MAR algorithm than with the conventional algorithm (96.7 ± 40.1 HU vs. 284.6 ± 134.1 HU, P < 0.001). Thirteen (27.7 %) patients were diagnosed as equivocal using the conventional algorithm but were diagnosed with perforation (2 patients) and non-perforation (11 patients) using the MAR algorithm (equivocal rate: 27.7 % vs. 0%, P < 0.001). Using the MAR algorithm, all cases were diagnosed with perforation (6 patients, 12.8 %) or non-perforation (41 patients, 87.2 %).

CONCLUSIONS: The MAR algorithm effectively reduced metal artifacts and allowed us to diagnose the presence or absence of perforation in all cases, whereas definitive diagnosis was difficult with the use of conventional algorithm in 27.7 % of cases.

来源:Eur J Radiol,查看原文。版权归原作者所有。