人工智能驱动的手持超声设备测量左心室射血分数
本研究纳入112例心内科住院患者,评估手持超声设备上LVivo软件自动测量左心室射血分数(LVEF)的准确性,以台式超声心动图三维测量为参考。LVivo与三维参考法的平均LVEF分别为46±14%和48±14%,相关性高(r=0.92,95%CI 0.87-0.95,p<0.0001),平均差-0.61%(95%CI -1.89至0.68,p=0.31)无显著差异。作者认为,在图像质量至少良好时,LVivo软件能够准确测量LVEF。
要点
- 手持超声设备LVivo自动测得的LVEF与三维超声参考法平均差-0.61%,无显著差异。
- 两者相关性高(r=0.92,95%CI 0.87-0.95),提示自动LVEF测量与三维参考一致性较好。
- 作者指出,采集图像质量至少良好时,LVivo能够准确测量LVEF。
关键影像表现/征象
- 四腔心尖切面是LVivo软件测量LVEF所用的切面。
诊断标准与关键数值
- LVivo与三维参考法平均LVEF分别为46±14%和48±14%。
- 两者相关性r=0.92(95%CI 0.87-0.95,p<0.0001)。
- 平均差-0.61%(95%CI -1.89至0.68,p=0.31),无显著差异。
临床应用与阅片要点
- 作者指出,采集图像质量至少良好时,LVivo软件能够准确测量LVEF。
局限性
- 原文提及LVivo软件存在局限性,但未具体说明。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The aim of this study was to assess the accuracy of an algorithm for automated measurement of left ventricular ejection fraction (LVEF) available on handheld ultrasound devices (HUDs). One hundred twelve patients admitted to the cardiology department underwent assessment performed with an HUD. In each case, the four-chamber apical view was obtained, and LVEF was calculated with LVivo software. Subsequently, during the examination performed with the use of the stationary echocardiograph, the 3-D measurement of LVEF was recorded. The average LVEFs measured with LVivo and the 3-D reference method were 46 ± 14% and 48 ± 14%, respectively. The correlation between the measurements obtained with the HUD and 3-D evaluation was high (r = 0.92, 95% confidence interval: 0.87-0.95, p < 0.0001). The mean difference between the LVEF obtained with LVivo and the 3-D LVEF was not significant (mean difference: -0.61%, 95% confidence interval: -1.89 to 0.68, p = 0.31). The LVivo software despite its limitations is capable of the accurate LVEF measurement when the acquired views are of at least good imaging quality.
引用
- Filipiak-Strzecka D, Kasprzak JD, Wejner-Mik P, et al. Artificial Intelligence-Powered Measurement of Left Ventricular Ejection Fraction Using a Handheld Ultrasound Device. Ultrasound Med Biol. 2021. PMID: 33451814. doi:10.1016/j.ultrasmedbio.2020.12.003
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。