儿科CT检查对Image Gently指南的临床依从性:53家临床机构663,417次CT扫描研究
该研究从53家机构获取663,417次腹部-盆腔和胸部CT扫描的剂量相关数据,按年龄队列比较儿童与成人队列的扫描参数和剂量指标。在150个儿科队列中,134个队列的扫描参数比成人更儿童化;其中128个队列的CTDIvol低于成人,111个队列的SSDE低于成人。研究重申Image Gently降低管电流和峰值千伏的建议在实践中被普遍采用并能有效降低儿科CT剂量。
要点
- 在150个儿科队列中,134个队列的扫描参数比成人更儿童化。
- 134个儿科队列中128个CTDIvol低于成人,128个中111个SSDE低于成人。
- Image Gently降低管电流和峰值千伏的建议在实践中常用且有效降低剂量。
核心结论
- 研究重申Image Gently降低管电流和峰值千伏的建议在实践中被普遍采用并能有效降低儿科CT剂量。
研究设计与样本
- 研究从53家机构获取663,417次腹部-盆腔和胸部CT扫描的剂量相关数据。
- 患者被分配至任意年龄队列,每组最少n=12例患者以满足统计需要。
- 所有儿科(<19岁)队列与成人队列用Kruskal-Wallis检验比较四个扫描参数:X线管千伏(kV)、管电流乘曝光时间乘积(管mAs)、扫描螺距和管旋转时间。
- 同样比较CTDIvol和SSDE以评估儿科队列是否接受更低剂量,P值<0.05视为显著。
诊断标准与关键数值
- 在150个儿科队列中,134个队列的扫描参数比成人更儿童化。
- 在134个儿科队列中,128个队列的CTDIvol低于成人。
- 在128个儿科队列中,111个队列的SSDE低于成人。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Managing patient radiation dose in pediatric computed tomography (CT) examinations is essential. Some organizations, most notably Image Gently, have suggested techniques to lower dose to pediatric patients and mitigate risk while maintaining image quality.
OBJECTIVE: We sought to validate whether institutions are observing Image Gently guidelines in practice.
MATERIALS AND METHODS: Dose-relevant data from 663,417 abdomen-pelvis and chest CT scans were obtained from 53 facilities. Patients were assigned arbitrary age cohorts with a minimum size of n=12 patients in each age group, for statistical purposes. All pediatric (<19 years old) cohorts at a given facility were compared to the adult cohort by a Kruskal-Wallis test for each of the four scan parameters - (1) x-ray tube kilovoltage (kV), (2) tube-current-by-exposure-time product (tube mAs), (3) scan pitch and (4) tube rotation time - to assess whether the distribution of values in the pediatric cohorts differed from the adult cohort. The same was repeated with volume CT dose index (CTDIvol) and size-specific dose estimate (SSDE) to assess whether pediatric cohorts received less dose than adult cohorts. A P-value of <0.05 was deemed significant.
RESULTS: Across the 150 pediatric cohorts, 134 had scan parameters that were more child-sized than their adult counterparts. In 128 of these 134 pediatric cohorts, the CTDIvol was less than the adult counterpart. In 111 of these 128 pediatric cohorts, the SSDE was less than the adult counterpart.
CONCLUSION: The study reaffirms that in practice, Image Gently's suggestions of lowering tube mAs and peak kilovoltage are commonly employed and effective at reducing pediatric CT dose.