文献速递胸部技术与安全DR

骨抑制软件在胸部X线摄影中提升图像质量并降低辐射剂量的价值

Value of bone suppression software in chest radiographs for improving image quality and reducing radiation dose.
PubMed 文献 · Eur Radiol · 2021-01-13

本研究对40例成人患者分别采用双能减影(DES)和骨抑制软件获取胸部X线骨抑制图像,比较图像质量指标与剂量面积乘积。软件法在骨区结构相似性更高但RMAE和PSNR更差,软组织区结构相似性更高,两位阅片者视觉评分均更高,且平均DAP显著低于DES法。

要点

  • 软件法骨抑制图像在骨区和软组织区的结构相似性指数均高于双能减影法。
  • 两位阅片者均给软件法骨抑制图像的视觉质量评分显著高于双能减影法。
  • 软件法平均剂量面积乘积为0.98 dGy·cm2,显著低于双能减影法的1.48 dGy·cm2。

关键影像表现/征象

  • 骨区:软件法骨抑制图像结构相似性指数优于双能减影法(94.57 vs. 87.77),但相对平均绝对误差和峰值信噪比更差(0.50 vs. 0.20;20.93 vs. 34.37),均p < 0.001。
  • 软组织区:软件法骨抑制图像结构相似性指数优于双能减影法(97.56 vs. 91.42),相对平均绝对误差和峰值信噪比与双能减影法相近(0.29 vs. 0.27;31.35 vs. 29.87),均p < 0.001。
  • 两位阅片者均给软件法骨抑制图像显著更高的图像质量评分(p < 0.001)。
  • 软件法相关图像平均剂量面积乘积为0.98 dGy·cm2,显著低于双能减影相关图像的1.48 dGy·cm2(p < 0.001)。

诊断标准与关键数值

  • 图像质量指标包括相对平均绝对误差、峰值信噪比和结构相似性指数,分别于骨区和软组织区计算。
  • 骨区结构相似性指数:软件法94.57 vs. 双能减影法87.77;相对平均绝对误差:软件法0.50 vs. 双能减影法0.20;峰值信噪比:软件法20.93 vs. 双能减影法34.37(均p < 0.001)。
  • 软组织区结构相似性指数:软件法97.56 vs. 双能减影法91.42;相对平均绝对误差:软件法0.29 vs. 双能减影法0.27;峰值信噪比:软件法31.35 vs. 双能减影法29.87(均p < 0.001)。
  • 剂量面积乘积:软件法0.98 dGy·cm2 vs. 双能减影法1.48 dGy·cm2(p < 0.001)。

临床应用与阅片要点

  • 骨抑制软件在保持软组织结构相似性方面优于双能减影技术。
  • 骨抑制软件在骨抑制图像中对肺部病变的图像质量优于双能减影技术。
  • 骨抑制软件较基于硬件的图像处理技术可降低辐射剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare image quality and radiation dose between dual-energy subtraction (DES)-based bone suppression images (D-BSIs) and software-based bone suppression images (S-BSIs).

METHODS: Chest radiographs (CXRs) of forty adult patients were obtained with the two X-ray devices, one with DES and one with bone suppression software. Three image quality metrics (relative mean absolute error (RMAE), peak signal-to-noise ratio (PSNR), and structural similarity index (SSIM)) between original CXR and BSI for each of D-BSI and S-SBI groups were calculated for each bone and soft tissue areas. Two readers rated the visual image quality for original CXR and BSI for each of D-BSI and S-SBI groups. The dose area product (DAP) values were recorded. Paired t test was used to compare the image quality and DAP values between D-BSI and S-BSI groups.

RESULTS: In bone areas, S-BSIs had better SSIM values than D-BSI (94.57 vs. 87.77) but worse RMAE and PSNR values (0.50 vs. 0.20; 20.93 vs. 34.37) (all p < 0.001). In soft tissue areas, S-BSIs had better SSIM values than D-BSI (97.56 vs. 91.42) but similar RMAE and PSNR values (0.29 vs. 0.27; 31.35 vs. 29.87) (all p < 0.001). Both readers gave S-BSIs significantly higher image quality scores than D-BSI (p < 0.001). The mean DAP in software-related images (0.98 dGy·cm2) was significantly lower than that in the DES-related images (1.48 dGy·cm2) (p < 0.001).

CONCLUSION: Bone suppression software significantly improved the image quality of bone suppression images with a relatively lower radiation dose, compared with dual-energy subtraction technique.

KEY POINTS: • Bone suppression software preserves structure similarity of soft tissues better than dual-energy subtraction technique in bone suppression images. • Bone suppression software achieves superior image quality for lung lesions than dual-energy subtraction technique in bone suppression images. • Bone suppression software can decrease the radiation dose over the hardware-based image processing technique.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Hong GS, Do KH, Son AY, et al. Value of bone suppression software in chest radiographs for improving image quality and reducing radiation dose. Eur Radiol. 2021. PMID: 33439320. doi:10.1007/s00330-020-07596-w

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。