5D血流MRI:一种完全自门控、自由运行的框架,用于心脏和呼吸运动分辨的3D血流动力学

5D Flow MRI: A Fully Self-gated, Free-running Framework for Cardiac and Respiratory Motion-resolved 3D Hemodynamics.
PubMed 文献 · Radiol Cardiothorac Imaging · 2020-11-12

前瞻性研究在体外搏动模体和20例成人中实现并验证了自门控自由运行全心脏5D血流MRI框架,可在7.65分钟±0.35内采集数据,与常规4D血流相比在升主动脉高估净流量和峰值速度(分别高达26%和12%),呼吸分辨分析显示下腔静脉吸气末净流量和峰值流量较呼气末显著增加。

要点

  • 5D血流MRI可在7.65分钟±0.35内完成全心脏采集,加速率36.0-76.9。
  • 与常规4D血流相比,升主动脉净流量和峰值速度分别高估高达26%和12%。
  • 呼吸分辨分析显示下腔静脉吸气末较呼气末净流量和峰值流量显著增加。

关键影像表现/征象

  • 体外5D血流结果与常规4D血流衍生值高度一致(峰值和净流量在所有量化平面上差异<7%)。
  • 体内5D血流与常规4D血流中度一致,升主动脉净流量和峰值速度分别高估高达26%和12%。
  • 体内5D血流在弓部和降主动脉低估<12%。
  • 呼吸分辨分析显示下腔静脉吸气末较呼气末净流量和峰值流量显著增加,上腔静脉呈相反趋势。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To implement, validate, and apply a self-gated free-running whole-heart five-dimensional (5D) flow MRI framework to evaluate respiration-driven effects on three-dimensional (3D) hemodynamics in a clinical setting.

MATERIALS AND METHODS: In this prospective study, a free-running five-dimensional (5D) flow sequence was implemented with 3D radial sampling, self-gating, and a compressed-sensing reconstruction. The 5D flow was evaluated in a pulsatile phantom and adult participants with aortic and/or valvular disease who were enrolled between May and August 2019. Conventional twofold-accelerated four-dimensional (4D) flow of the thoracic aorta with navigator gating was performed as a reference comparison. Continuous parameters were evaluated for parameter normality and were compared between conventional 4D flow and 5D flow using a signed-rank or two-tailed paired t test. Differences between respiratory states were evaluated using a repeated-measure analysis of variance or a nonparametric Friedman test.

RESULTS: A total of 20 adult participants (mean age, 49 years ± 17 [standard deviation]; 18 men and two women) were included. In vitro 5D flow results showed excellent agreement with conventional 4D flow-derived values (peak and net flow, <7% difference over all quantified planes). Whole-heart 5D flow data were collected in all participants in 7.65 minutes ± 0.35 (acceleration rate = 36.0-76.9) versus 9.88 minutes ± 3.17 for conventional aortic 4D flow. In vivo, 5D flow demonstrated moderate agreement with conventional 4D flow but demonstrated overestimation in net flow and peak velocity (up to 26% and 12%, respectively) in the ascending aorta and underestimation (<12%) in the arch and descending aorta. Respiratory-resolved analyses of caval veins showed significantly increased net and peak flow in the inferior vena cava in end inspiration compared with end expiration, and the opposite trend was shown in the superior vena cava.

CONCLUSION: A free-running 5D flow MRI framework consistently captured cardiac and respiratory motion-resolved 3D hemodynamics in less than 8 minutes. Supplemental material is available for this article. © RSNA, 2020.