使用有效直径对体重未知的成人患者进行CT剂量管理
本研究回顾性分析642例成人患者,评估有效直径(Deff)用于体重未知成人CT剂量管理的价值。结果显示体重与Deff的相关系数在胸部CT中为0.920和0.929,腹部CT为0.805;体重50-70kg组与Deff等效于体重50-70kg组的容积CT剂量指数和剂量长度乘积均无显著差异。结论认为Deff可能作为体重未知成人CT剂量管理的体型参数。
要点
- 体重与Deff的相关系数:胸部CT(Aquilion Prime SP)0.920,胸部CT(Biograph mCT)0.929,腹部CT(Aquilion Prime SP)0.805。
- 体重50-70kg组与Deff等效于体重50-70kg组在胸部和腹部CT中的容积CT剂量指数和剂量长度乘积均无显著差异。
- Deff由剂量管理软件(Radimetrics)从CT定位像上的侧方直径自动计算。
核心结论
- Deff可能作为体重未知成人患者CT剂量管理的体型参数。
研究设计与样本
- 回顾性研究,纳入2018年4月至2019年9月间642例CT检查前已测量身高和体重的成人患者。
- 胸部CT使用Aquilion Prime SP 428例,胸部CT使用Biograph mCT 100例,腹部CT使用Aquilion Prime SP 114例。
- Deff由剂量管理软件(Radimetrics)从CT定位像上的侧方直径自动计算。
- 采用Pearson积矩相关分析体重与Deff的相关性,采用t检验进行统计分析。
诊断标准与关键数值
- 体重与Deff的相关系数:胸部CT(Aquilion Prime SP)为0.920,胸部CT(Biograph mCT)为0.929,腹部CT(Aquilion Prime SP)为0.805。
- 胸部和腹部CT中,体重50-70kg组与Deff等效于体重50-70kg组的容积CT剂量指数和剂量长度乘积均无显著差异。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the usefulness of effective diameter (Deff) for CT dose management of adult patients with unknown body weight.
METHODS: A total of 642 adult patients whose height and weight had been measured before CT examination (chest CT using Aquilion Prime SP, 428 patients; chest CT using Biograph mCT, 100 patients; and abdominal CT using Aquilion Prime SP, 114 patients) were retrospectively examined between April 2018 and September 2019. The Deff was automatically calculated from the lateral diameter on a CT localizer radiograph by a dose management software (Radimetrics). In order to determine the correlation between body weight and Deff, we compared volume CT dose index and dose length product between patients with body weight between 50 and 70 kg and those with Deff equivalent to body weight between 50 and 70 kg. Correlation analysis was performed by Pearson's product-moment correlation, and statistical analyses were performed by using t-test.
RESULTS: The correlation coefficient values between body weight and Deff were 0.920 for chest CT using Aquilion Prime SP, 0.929 for chest CT using Biograph mCT, and 0.805 for abdominal CT using Aquilion Prime SP. In both chest and abdominal CT scans, there were no significant differences in volume CT dose index and dose length product between patients with body weight between 50 and 70 kg and those with Deff equivalent to body weight between 50 and 70 kg.
CONCLUSIONS: The Deff may be useful as a somatometric parameter for CT dose management of adult patients with unknown body weight.
引用
- Fukunaga M, Matsubara K, Ichikawa S, et al. CT dose management of adult patients with unknown body weight using an effective diameter. Eur J Radiol. 2020. PMID: 33388531. doi:10.1016/j.ejrad.2020.109483
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。