X线、计算机断层扫描与磁共振成像对外来异物材料的可检测性:体模研究
研究将34种常见异物(含木材、塑料、玻璃等)嵌入明胶体模,比较X线、CT和MR的显示能力。MR检测敏感度最高(97.1%),CT次之(86.0%),X线最低(61.8%);正确识别异物及材料类型的成功率也以MR最高(33.3%和39.2%)。
要点
- MR对非铁磁性异物的检测敏感度最高(97.1%),CT次之(86.0%),X线最低(61.8%)。
- 正确识别异物及材料类型的成功率MR最高(33.3%和39.2%),CT均为25.5%,X线为16.7%和15.7%。
- 主观可见性CT和MR评为「可见度良好」,X线评为「可见度差」。
诊断标准与关键数值
- 异物检测敏感度:MR 97.1%,CT 86.0%,X线 61.8%。
- 正确识别物体的成功率:MR 33.3%,CT 25.5%,X线 16.7%。
- 正确识别材料类型的成功率:MR 39.2%,CT 25.5%,X线 15.7%。
- 主观可见性:CT和MR评为「可见度良好」,X线评为「可见度差」。
- 读者间一致性高:Kendall's W分别为X线0.935、MR 0.834、CT 0.794。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the effectiveness of plain radiography (X-ray. XR), computed tomography (CT) and magnetic resonance imaging (MR) in visualising commonly seen foreign bodies. A special focus was put on objects relevant to head and neck surgery.
METHOD: Thirty-four commonly encountered objects of different compositions including wood, plastic, and glass were embedded in a gelatin gel phantom and imaged using XR, CT and MR. The success rates of radiologists in detecting and correctly identifying the foreign objects were evaluated. Subjective visibility was rated on a 4-point Likert scale. Objective visibility was analysed using region of interest-based contrast for CT.
RESULTS: Sensitivity in foreign bodies detection was highest in MR (97.1 %) followed by CT (86.0 %) and x-ray (61.8 %). Success rates for the correct identification of the objects and material types were highest in MR (33.3 % and 39.2 %, respectively) followed by CT (25.5 % for both) and XR (16.7 % and 15.7 %). Overall, subjective visibility was rated higher in CT and MR imaging ("good visibility"), as compared to XR ("poor visibility"). Interreader agreement was high across modalities (Kendall's W = 0.935, 0.834 and 0.794 for XR, MR and CT, respectively).
CONCLUSIONS: Detection and identification of non-ferromagnetic objects was most successful in MR followed by CT imaging in this experimental setup.
引用
- Voss JO, Doll C, Raguse JD, et al. Detectability of foreign body materials using X-ray, computed tomography and magnetic resonance imaging: A phantom study. Eur J Radiol. 2020. PMID: 33421828. doi:10.1016/j.ejrad.2020.109505