压缩感知与并行成像加速T2 FSE序列在头颈部MR成像中的比较:常规临床实践中的效用评估
本研究纳入30例连续头颈部疾病患者,对比并行成像(PI)联合压缩感知(CS)与单纯PI加速的T2 FSE序列,发现CS组平均检查时间83.5秒显著短于PI组173.0秒,且信噪比和对比噪声比均显著更优,定性评估一致性良好,结论为该序列在缩短扫描时间的同时图像质量相当或略优。
要点
- CS组信噪比(11.2±3.6)和对比噪声比中位数(7.4)均显著优于PI组(8.9±2.6;6.1)。
- 两名评估者对所有定性指标的一致性为显著且基本一致(0.62<κ<0.81)。
核心结论
- 在头颈部MR成像中,PI联合CS加速的T2加权序列在缩短扫描时间的情况下,表现与常规PI加速序列相当甚至略优。
研究设计与样本
- 研究为连续纳入30例(男15例、女15例,平均年龄53±22岁)各种头颈部疾病患者的影像学对比研究。
- 所有患者均接受PI联合CS(缩减因子3)和单纯PI(缩减因子1.5)的T2 FSE序列MR成像。
- 定量评估计算信噪比和对比噪声比;定性评估由两名研究者采用5分制对整体图像质量、伪影和诊断信心水平评分,最终评分由两名读者共识确定。
- 统计方法采用配对t检验和Wilcoxon符号秩检验比较平均检查时间和所有指标,采用kappa统计评估观察者间一致性。
关键影像表现/征象
- PI联合CS序列的平均检查时间为83.5±11.0秒,显著短于单纯PI序列的173.0±54.4秒(p<0.0001)。
- PI联合CS序列的平均信噪比为11.2±3.6,显著优于单纯PI序列的8.9±2.6(p<0.0001)。
- PI联合CS序列的对比噪声比中位数为7.4,显著优于单纯PI序列的6.1(p<0.0001)。
诊断标准与关键数值
- 定性评估采用5分制评分系统,评估整体图像质量、伪影和诊断信心水平,最终评分由两名读者共识确定。
- 观察者间一致性采用kappa统计评估,所有定性指标的一致性均为显著且基本一致(0.62<κ<0.81)。
临床应用与阅片要点
- PI联合CS加速的T2 FSE序列可在头颈部MR常规临床实践中缩短检查时间,同时保持或略优于常规PI序列的图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To directly compare the capability of compressed sensing (CS) and parallel imaging (PI) accelerated T2 FSE (Fast Spin Echo) sequence with PI for head and neck MR imaging.
METHODS: Thirty consecutive patients with various head and neck diseases (15 men and 15 women, mean age 53 ± 22 years) underwent MR imaging by PI with CS and by PI. Reduction factors were as follows: PI with CS, 3 and PI, 1.5. Examination times for PI with CS and PI were all recorded. For quantitative image quality assessment, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. For qualitative assessment, two investigators assessed overall image quality, artifacts and diagnostic confidence level using a 5-point scoring system, and final scores were determined by consensus of two readers. Mean examination time and all indexes were compared by means of paired t-test and Wilcoxon signed-rank test. Inter-observer agreement for each qualitative index was assessed in terms of kappa statistics.
RESULTS: Mean examination time for PI with CS (83.5 ± 11.0 s) was significantly shorter than that for PI (173.0 ± 54.4 s, p < 0.0001). SNR and CNR of PI with CS were significantly better than those with PI (mean SNR; 11.2 ± 3.6 vs 8.9 ± 2.6, median of CNR; 7.4 vs. 6.1, p < 0.0001). All inter-observer agreements were assessed as significant and substantial (0.62 < κ < 0.81).
CONCLUSION: PI with CS accelerated T2 weighted sequence performs equally well or even slightly better than its PI accelerated, conventional counterpart at reduced scan times in the context of head and neck MR imaging.
引用
- Ikeda H, Ohno Y, Murayama K, et al. Compressed sensing and parallel imaging accelerated T2 FSE sequence for head and neck MR imaging: Comparison of its utility in routine clinical practice. Eur J Radiol. 2020. PMID: 33395594. doi:10.1016/j.ejrad.2020.109501
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。