锥形束CT与多探测器CT在人工耳蜗植入术后成像中的比较:图像质量与辐射剂量的评估
这项回顾性多中心研究纳入51例人工耳蜗植入患者,比较术后颞骨锥形束CT(32耳)与多探测器CT(19耳)对植入物位置及精细解剖结构的显示质量、金属伪影和辐射剂量。锥形束CT在显示电极阵列位置、耳蜗外壁、单个电极触点及骨螺旋板方面更敏感,金属伪影更少,剂量长度乘积显著更低;多探测器CT在骨和空气中的信噪比更高。
要点
- 锥形束CT显示单个电极触点、电极阵列位置、耳蜗外壁及骨螺旋板优于多探测器CT,金属伪影更少。
- 两种方法测量电极植入角度的能力无显著差异,显示蜗轴、耳蜗内壁及乳突段面神经管壁也无显著差异。
- 锥形束CT剂量长度乘积显著低于多探测器CT,但多探测器CT在骨和空气中的信噪比更高。
关键影像表现/征象
- 锥形束CT显示电极阵列位置更敏感(P=0.046),显示耳蜗外壁更敏感(P=0.001),显示单个电极触点更敏感(P<0.001),显示骨螺旋板更敏感(P=0.004),金属伪影更少(P<0.001)。
- 两种方法显示蜗轴(P=0.37)、耳蜗内壁(P>0.99)及乳突段面神经管壁(P=0.07)无显著差异,测量电极植入角度的能力也无显著差异(P>0.99)。
- 锥形束CT组剂量长度乘积显著更低(P<0.001),多探测器CT在骨中(P=0.22)和空气中(P=0.001)的信噪比更高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Cone-beam CT is being increasingly used in head and neck imaging. We compared cone-beam CT with multidetector CT to assess postoperative implant placement and delineate finer anatomic structures, image quality, and radiation dose used.
MATERIALS AND METHODS: This retrospective multicenter study included 51 patients with cochlear implants and postoperative imaging via temporal bone cone-beam CT (n = 32 ears) or multidetector CT (n = 19 ears) between 2012 and 2017. We evaluated the visualization quality of single electrode contacts, the scalar position of the electrodes, cochlear walls, mastoid facial canal, metallic artifacts (using a 4-level visual score), and the ability to measure the insertion angle of the electrodes. The signal-to-noise ratio and radiation dose were also evaluated.
RESULTS: Cone-beam CT was more sensitive for visualizing the scalar position of the electrodes (P = .046), cochlear outer wall (P = .001), single electrode contacts (P < .001), and osseous spiral lamina (P = .004) and had fewer metallic artifacts (P < .001). However, there were no significant differences between both methods in visualization of the modiolus (P = .37), cochlear inner wall (P > .99), and mastoid facial canal wall (P = .07) and the ability to measure the insertion angle of the electrodes (P > .99). The conebeam CT group had significantly lower dose-length product (P < .001), but multidetector CT showed a higher signal-to-noise ratio in both bone and air (P = .22 and P = .001).
CONCLUSIONS: Cone-beam CT in patients with cochlear implants provides images with higher spatial resolution and fewer metallic artifacts than multidetector CT at a relatively lower radiation dose.
引用
- Helal RA, Jacob R, Elshinnawy MA, et al. Cone-beam CT versus Multidetector CT in Postoperative Cochlear Implant Imaging: Evaluation of Image Quality and Radiation Dose. AJNR Am J Neuroradiol. 2021. PMID: 33414229. doi:10.3174/ajnr.A6894
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。