人工耳蜗电极阵列植入后颞骨标本的影像-组织学对比研究:锥形束CT是否优于多层螺旋CT?
本研究回顾性纳入20例植入电极后的颞骨标本,由两名放射科医生独立评估锥形束CT(96 kV与120 kV)和多层螺旋CT的图像质量及电极位置诊断准确性,并与组织学检查对照。结果显示锥形束CT在多数骨结构显示和电极接触点可视化上显著优于多层螺旋CT,且120 kV锥形束CT诊断电极腔内位置的准确率最高(100%)。
要点
- 锥形束CT在耳囊骨结构、耳蜗壁、骨螺旋板及电极位置显示上显著优于多层螺旋CT。
- 120 kV与96 kV锥形束CT间无显著差异,但120 kV诊断电极腔内位置的准确率更高(100% vs 97.5%)。
- 多层螺旋CT仅正确诊断77.5%的标本电极腔内位置,低于锥形束CT。
核心结论
- 数据提示,在人工耳蜗植入后颞骨标本成像中,锥形束CT的诊断准确性高于多层螺旋CT,尤其用于确定植入物的耳蜗内定位。
研究设计与样本
- 回顾性研究,纳入20例电极植入后的颞骨标本,由两名放射科医生独立评估锥形束CT(96 kV和120 kV)与多层螺旋CT图像,以组织学检查为参考标准。
关键影像表现/征象
- 锥形束CT对耳囊骨结构、耳蜗壁(除外外部)、骨螺旋板、电极位置、阵列上单个电极接触点、金属伪影及整体图像质量的评分显著高于多层螺旋CT(最大P=.04)。
- 锥形束CT 96 kV与120 kV之间各评估项目无显著差异(最小P=.21)。
诊断标准与关键数值
- 采用五点量表定性评估骨结构、电极位置、电极接触点显示、金属伪影和整体图像质量。
- 与组织学对照,120 kV锥形束CT正确诊断100%标本的耳蜗内电极位置,96 kV锥形束CT为97.5%,多层螺旋CT为77.5%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE. The purpose of this article is to evaluate subjective image quality and diagnostic accuracy to determine cochlear implant (CI) electrode position in a temporal bone (TB) specimen on cone-beam CT (CBCT) versus MDCT. MATERIALS AND METHODS. In this retrospective study, two radiologists independently reviewed CBCT (96-kV and 120-kV settings) and MDCT images of 20 TB specimens after electrode implantation. Qualitative evaluation of bone structures of the otic capsule, inner and outer cochlear wall, osseous spiral lamina, electrode position relative to the osseous spiral lamina, visualization of single electrode contacts on the array, metal artifacts, and overall image quality was performed using a five-point scale. Intracochlear electrode position was subsequently correlated with histologic examination. RESULTS. Radiologic assessment of bone structures of the otic capsule, the cochlear wall (except the outer part), osseous spiral lamina, electrode position, visualization of single electrode contacts on the array, metal artifacts, and overall image quality were significantly higher in CBCT compared with MDCT (maximum p = .04). No significant differences were found between CBCT at 96 kV and 120 kV (minimum p = .21). The intracochlear electrode position with histologic correlation was correctly diagnosed in 100% and 97.5% of specimens on 120-kV and 96-kV CBCT, respectively, whereas 77.5% were correctly assessed using MDCT. CONCLUSION. The data suggest that CBCT shows a higher diagnostic accuracy in TB specimen imaging after CI compared with MDCT, in particular to determine the intracochlear localization of the implant.
引用
- Burck I, Schneider SV, Balster S, et al. Radiohistologic Comparison Study of Temporal Bone Specimens After Cochlear Implant Electrode Array Insertion: Is Cone-Beam CT Superior to MDCT? AJR Am J Roentgenol. 2021. PMID: 33439050. doi:10.2214/AJR.20.23157