采用策略性采集梯度回波(STAGE)的定量MRI:针对1.5 T扫描仪的优化及T1弛豫图验证
本研究将原为3T扫描仪开发的STAGE序列适配至1.5T,通过理论模拟、体模和活体采集优化翻转角,并评估STAGE衍生T1值的准确性与可重复性。优化的STAGE方案可在1.5T上约7分钟内提供准确的T1图及其他定性图像和定量图。
要点
- 1.5T下STAGE最优翻转角对为FA1,2=7°、38°。
- STAGE衍生T1估计误差在体模中低于3%、活体中低于5%。
- 优化STAGE方案在1.5T上约7分钟可同时获得T1、T2*、质子密度和定量磁化率图。
关键影像表现/征象
- STAGE序列原为3T扫描仪开发,可在约5分钟内衍生T1、T2*、质子密度和定量磁化率图。
- 针对1.5T扫描仪适配STAGE序列,并通过理论模拟、体模和活体采集优化翻转角。
诊断标准与关键数值
- 1.5T下最优翻转角对为FA1,2=7°、38°。
- STAGE衍生T1估计误差在体模中低于3%,在活体测量中低于5%。
- 不同组织的T1变异系数(COV)在生理变异性方面为1.8%至4.8%,在扫描-重扫描可重复性方面为0.8%至2%。
临床应用与阅片要点
- 作者指出,优化的STAGE方案可在1.5T扫描仪上约7分钟内提供准确的T1图及其他定性图像和定量图。
- 作者指出,本研究为评估STAGE在临床环境中的作用提供了基础。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The strategically acquired gradient echo (STAGE) protocol, developed for 3T scanners, allows one to derive quantitative maps such as T1, T2*, proton density, and quantitative susceptibility mapping in about 5 min. Our aim was to adapt the STAGE sequences for 1.5T scanners which are still commonly used in clinical practice. Furthermore, the accuracy and repeatability of the STAGE-derived T1 estimate were tested.
METHODS: Flip angle (FA) optimization was performed using a theoretical simulation by maximizing signal-to-noise ratio, contrast-to-noise ratio, and T1 precision. The FA choice was further refined with the ISMRM/NIST phantom and in vivo acquisitions. The accuracy of the T1 estimate was assessed by comparing STAGE-derived T1 values with T1 maps obtained with an inversion recovery sequence. T1 accuracy was investigated for both the phantom and in vivo data. Finally, one subject was acquired 10 times once a week and a group of 27 subjects was scanned once. The T1 coefficient of variation (COV) was computed to assess scan-rescan and physiological variability, respectively.
RESULTS: The FA1,2 = 7°,38° were identified as the optimal FA pair at 1.5T. The T1 estimate errors were below 3% and 5% for phantom and in vivo measurements, respectively. COV for different tissues ranged from 1.8 to 4.8% for physiological variability, and between 0.8 and 2% for scan-rescan repeatability.
CONCLUSION: The optimized STAGE protocol can provide accurate and repeatable T1 mapping along with other qualitative images and quantitative maps in about 7 min on 1.5T scanners. This study provides the groundwork to assess the role of STAGE in clinical settings.
KEY POINTS: • The STAGE imaging protocol was optimized for use on 1.5T field strength scanners. • A practical STAGE protocol makes it possible to derive quantitative maps (i.e., T1, T2*, PD, and QSM) in about 7 min at 1.5T. • The T1 estimate derived from the STAGE protocol showed good accuracy and repeatability.