文献速递神经介入DSA

使用数据驱动模型与定量血管造影方法评估血栓切除术中的再灌注状态(mTICI)

Use of quantitative angiographic methods with a data-driven model to evaluate reperfusion status (mTICI) during thrombectomy.
PubMed 文献 · Neuroradiology · 2021-01-07

本研究基于383例前循环大血管闭塞行机械取栓患者的DSA,将晚期动脉期与毛细血管期图像量化为血管造影参数成像(API)峰值高度图并归一化,用卷积神经网络分类再灌注状态。二分类(mTICI 0,1,2a vs 2b,2c,3)平均准确率81.0%(95%CI 79.0-82.9%),AUROC 0.86(0.84-0.88);三分类平均准确率64.0%(62.0-66.0)。结论认为该数据驱动方法可自动评估术中再灌注,具有可行性。

要点

  • API峰值高度图联合CNN可在术中自动分类mTICI再灌注状态。
  • 二分类平均准确率81.0%(95%CI 79.0-82.9%),AUROC 0.86(0.84-0.88)。
  • 三分类平均准确率64.0%(62.0-66.0),mTICI 0,1,2a类AUROC 0.85(0.83-0.87)。

关键影像表现/征象

  • 利用API峰值高度图可自动分类mTICI再灌注状态。

诊断标准与关键数值

  • 二分类(mTICI 0,1,2a vs 2b,2c,3)平均准确率81.0%(95%CI 79.0-82.9%),AUROC 0.86(0.84-0.88)。
  • 三分类平均准确率64.0%(62.0-66.0)。
  • 三分类中mTICI 0,1,2a类AUROC 0.85(0.83-0.87)、mTICI 2b类AUROC 0.74(0.71-0.77)、mTICI 2c,3类AUROC 0.78(0.76-0.81)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Intra-procedural assessment of reperfusion during mechanical thrombectomy (MT) for emergent large vessel occlusion (LVO) stroke is traditionally based on subjective evaluation of digital subtraction angiography (DSA). However, semi-quantitative diagnostic tools which encode hemodynamic properties in DSAs, such as angiographic parametric imaging (API), exist and may be used for evaluation of reperfusion during MT. The objective of this study was to use data-driven approaches, such as convolutional neural networks (CNNs) with API maps, to automatically assess reperfusion in the neuro-vasculature during MT procedures based on the modified thrombolysis in cerebral infarction (mTICI) scale.

METHODS: DSAs from patients undergoing MTs of anterior circulation LVOs were collected, temporally cropped to isolate late arterial and capillary phases, and quantified using API peak height (PH) maps. PH maps were normalized to reduce injection variability. A CNN was developed, trained, and tested to classify PH maps into 2 outcomes (mTICI 0,1,2a/mTICI 2b,2c,3) or 3 outcomes (mTICI 0,1,2a/mTICI 2b/mTICI 2c,3), respectively. Ensembled networks were used to combine information from multiple views (anteroposterior and lateral).

RESULTS: The study included 383 DSAs. For the 2-outcome classification, average accuracy was 81.0% (95% CI, 79.0-82.9%), and the area under the receiver operating characteristic curve (AUROC) was 0.86 (0.84-0.88). For the 3-outcome classification, average accuracy was 64.0% (62.0-66.0), and AUROC values were 0.85 (0.83-0.87), 0.74 (0.71-0.77), and 0.78 (0.76-0.81) for the mTICI 0,1,2a, mTICI 2b, and mTICI 2c,3 classes, respectively.

CONCLUSION: This study demonstrated the feasibility of using hemodynamic information in API maps with data-driven models to autonomously assess intra-procedural reperfusion during MT.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Shiraz Bhurwani MM, Snyder KV, Waqas M, et al. Use of quantitative angiographic methods with a data-driven model to evaluate reperfusion status (mTICI) during thrombectomy. Neuroradiology. 2021. PMID: 33415348. doi:10.1007/s00234-020-02598-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。