文献速递神经CT

人工智能识别大血管闭塞在综合卒中中心的评估

Evaluation of Artificial Intelligence-Powered Identification of Large-Vessel Occlusions in a Comprehensive Stroke Center.
PubMed 文献 · AJNR Am J Neuroradiol · 2020-12-31

在综合卒中中心连续纳入2018年1月至2019年3月的1167例头颈部CTA,评估Viz LVO算法检测大血管闭塞的准确性,以资深神经放射学家的正式报告为参考标准。全部CTA中6.4%存在大血管闭塞,系统敏感度0.81、阴性预测值0.99、准确率0.94;卒中方案亚组敏感度0.82、阴性预测值0.96、准确率0.89。

要点

  • 全部1167例头颈部CTA中,系统检测大血管闭塞的敏感度为0.81、阴性预测值为0.99、准确率为0.94。
  • 卒中方案亚组404例中,系统检测大血管闭塞的敏感度为0.82、阴性预测值为0.96、准确率为0.89。
  • 作者认为该系统有潜力早期识别大血管闭塞卒中患者,希望改善未来管理与卒中救治。

核心结论

  • 作者认为Viz LVO在真实临床实践中具有早期识别大血管闭塞卒中患者的潜力,希望改善未来管理与卒中救治。

研究设计与样本

  • 本研究在综合卒中中心安装Viz LVO,对2018年1月至2019年3月所有连续头颈部CTA进行算法扫描以检测大血管闭塞。
  • 以资深神经放射学家的正式报告作为大血管闭塞存在的参考标准,将系统结果与参考标准比较。
  • 共纳入1167例头颈部CTA,其中404例为卒中方案。

诊断标准与关键数值

  • 全部1167例CTA中,参考标准确认75例(6.4%)存在大血管闭塞,系统检测出61例,敏感度0.81、阴性预测值0.99、准确率0.94。
  • 卒中方案亚组404例中,72例(17.8%)存在大血管闭塞,系统识别59例,敏感度0.82、阴性预测值0.96、准确率0.89。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Artificial intelligence algorithms have the potential to become an important diagnostic tool to optimize stroke workflow. Viz LVO is a medical product leveraging a convolutional neural network designed to detect large-vessel occlusions on CTA scans and notify the treatment team within minutes via a dedicated mobile application. We aimed to evaluate the detection accuracy of the Viz LVO in real clinical practice at a comprehensive stroke center.

MATERIALS AND METHODS: Viz LVO was installed for this study in a comprehensive stroke center. All consecutive head and neck CTAs performed from January 2018 to March 2019 were scanned by the algorithm for detection of large-vessel occlusions. The system results were compared with the formal reports of senior neuroradiologists used as ground truth for the presence of a large-vessel occlusion.

RESULTS: A total of 1167 CTAs were included in the study. Of these, 404 were stroke protocols. Seventy-five (6.4%) patients had a large-vessel occlusion as ground truth; 61 were detected by the system. Sensitivity was 0.81, negative predictive value was 0.99, and accuracy was 0.94. In the stroke protocol subgroup, 72 (17.8%) of 404 patients had a large-vessel occlusion, with 59 identified by the system, showing a sensitivity of 0.82, negative predictive value of 0.96, and accuracy of 0.89.

CONCLUSIONS: Our experience evaluating Viz LVO shows that the system has the potential for early identification of patients with stroke with large-vessel occlusions, hopefully improving future management and stroke care.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Yahav-Dovrat A, Saban M, Merhav G, et al. Evaluation of Artificial Intelligence-Powered Identification of Large-Vessel Occlusions in a Comprehensive Stroke Center. AJNR Am J Neuroradiol. 2020. PMID: 33384294. doi:10.3174/ajnr.A6923

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。