文献速递技术与安全外周血管US

使用全变分正则化反卷积改进超声微血管成像

Improved Ultrasound Microvessel Imaging Using Deconvolution with Total Variation Regularization.
PubMed 文献 · Ultrasound Med Biol · 2021-01-16

本研究提出一种全变分正则化反卷积方法,用于提高超声微血管成像的空间分辨率并减轻振铃伪影。在鸡胚脑、离体鸡胚绒毛尿囊膜和肿瘤数据上评估,重建的功率多普勒图像空间分辨率显著优于原始图像:微血管半高全宽从132 µm改善至83 µm,223条血管的平均半高全宽从233.19 ± 85.08 µm降至172.31 ± 75.11 µm(p < 0.0001),对比噪声比从1.06 dB提高到4.03 dB。与Tikhonov正则化相比,所提方法未出现假微血管,表现更优。

要点

  • 全变分正则化反卷积可显著提高功率多普勒微血管图像的空间分辨率,半高全宽从132 µm改善至83 µm。
  • 223条血管的平均半高全宽从233.19 ± 85.08 µm降至172.31 ± 75.11 µm,差异有统计学意义(p < 0.0001)。
  • 对比噪声比从1.06 dB提高到4.03 dB,且未出现Tikhonov正则化中的假微血管。

关键影像表现/征象

  • 重建的功率多普勒图像空间分辨率显著提高:微血管横截面轮廓的半高全宽从132 µm改善至83 µm。
  • 相距154 µm的两条相邻微血管使用所提方法比常规功率多普勒成像分离得更好。
  • 223条血管的平均半高全宽从233.19 ± 85.08 µm降至172.31 ± 75.11 µm,最大半高全宽从693.01 µm降至668.69 µm,最小半高全宽从73.92 µm降至45.74 µm。
  • 对比噪声比从1.06 dB提高到4.03 dB。
  • 与Tikhonov正则化相比,Tikhonov正则化出现假微血管而所提方法未出现。

诊断标准与关键数值

  • 基于223条血管横截面轮廓的半高全宽比较,所提方法与原始方法之间差异有统计学意义(秩和检验,p < 0.0001)。
  • 空间分辨率以半高全宽衡量:单条微血管从132 µm改善至83 µm;223条血管平均从233.19 ± 85.08 µm降至172.31 ± 75.11 µm。
  • 对比噪声比从1.06 dB提高到4.03 dB。

临床应用与阅片要点

  • 所提方法可能为功率多普勒微血管成像提供更高空间分辨率和对比噪声比,并减少Tikhonov正则化相关的假微血管伪影。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Singular value decomposition-based clutter filters can robustly reject tissue clutter, allowing for detection of slow blood flow in imaging microvasculature. However, to identify microvessels, high ultrasound frequency must be used to increase the spatial resolution at the expense of shorter depth of penetration. Deconvolution using Tikhonov regularization is an imaging processing method widely used to improve spatial resolution. The ringing artifact of Tikhonov regularization, though, can produce image artifacts such as non-existent microvessels, which degrade image quality. Therefore, a deconvolution method using total variation is proposed in this study to improve spatial resolution and mitigate the ringing artifact. Performance of the proposed method was evaluated using chicken embryo brain, ex ovo chicken embryo chorioallantoic membrane and tumor data. Results revealed that the reconstructed power Doppler (PD) images are substantially improved in spatial resolution compared with original PD images: the full width half-maximum (FWHM) of the cross-sectional profile of a microvessel was improved from 132 to 83 µm. Two neighboring microvessels that were 154 µm apart were better separated using the proposed method than conventional PD imaging. Additionally, 223 FWHMs measured from the cross-sectional profiles of 223 vessels were used to determine the improvement in FWHM with the proposed method statistically. The mean ± standard deviation of the FWHM without and with the proposed method was 233.19 ± 85.08 and 172.31 ± 75.11 μm, respectively; the maximum FWHM without and with the proposed method was 693.01 and 668.69 μm; and the minimum FWHM without and with the proposed method was 73.92 and 45.74 μm. There were statistically significant differences between FWHMs with and without the proposed method according to the rank-sum test, p < 0.0001. The contrast-to-noise ratio improved from 1.06 to 4.03 dB with use of the proposed method. We also compared the proposed method with Tikhonov regularization using ex ovo chicken embryo chorioallantoic membrane data. We found that the proposed method outperformed Tikhonov regularization as false microvessels appeared using the Tikhonov regularization but not with the proposed method. These results indicate that the proposed method is capable of providing more robust PD images with higher spatial resolution and higher contrast-to-noise ratio.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Lok UW, Trzasko JD, Huang C, et al. Improved Ultrasound Microvessel Imaging Using Deconvolution with Total Variation Regularization. Ultrasound Med Biol. 2021. PMID: 33468358. doi:10.1016/j.ultrasmedbio.2020.12.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。