基于MRI的合成CT在疑似骶髂关节炎患者结构病变检测中的应用:与MRI的比较
前瞻性研究纳入30例临床怀疑炎性骶髂关节炎的成人,同日行骶髂关节MRI和CT,由MRI生成合成CT(sCT),以CT共识结果为参考标准比较sCT与T1加权MRI对侵蚀、硬化和关节强直的诊断效能。sCT对侵蚀、硬化和关节强直的诊断准确率均高于T1加权MRI,其中侵蚀检测特异度、硬化检测敏感度和关节强直检测敏感度均得到提高。
要点
- sCT对骶髂关节侵蚀、硬化和关节强直的诊断准确率均高于常规T1加权MRI。
- sCT检测侵蚀的特异度为96%,高于T1加权MRI的89%。
- sCT检测硬化和关节强直的敏感度分别为94%和93%,均高于T1加权MRI。
核心结论
- 以CT为参考标准,基于MRI的骶髂关节合成CT在检测疑似骶髂关节炎患者的结构病变方面,诊断效能优于常规T1加权MRI。
研究设计与样本
- 前瞻性研究(临床试验注册号B670201837885),2019年2月至2019年11月纳入30例来自三级医院风湿科门诊、临床怀疑炎性骶髂关节炎的成人。
- 同日行骶髂关节MRI和CT检查,sCT图像由MRI扫描使用市售深度学习图像合成方法生成。
- 两名阅片者分别在不同阅片会话中独立记录T1加权MRI、sCT和CT上的结构病变(侵蚀、硬化和关节强直),阅片者对临床信息和其他图像设盲。
- 以CT共识结果为参考标准,采用广义估计方程模型分析sCT和T1加权MRI的诊断效能。
关键影像表现/征象
- sCT与T1加权MRI均可显示骶髂关节的侵蚀、硬化和关节强直,sCT对上述结构病变的检出准确率更高。
诊断标准与关键数值
- sCT诊断侵蚀的准确率为94%,T1加权MRI为86%(P = .003)。
- sCT诊断硬化的准确率为97%,T1加权MRI为81%(P < .001)。
- sCT诊断关节强直的准确率为92%,T1加权MRI为84%(P = .04)。
- sCT检测侵蚀的特异度为96%(95% CI: 90, 98),T1加权MRI为89%(95% CI: 81, 94)(P = .01)。
- sCT检测硬化的敏感度为94%(95% CI: 87, 97),T1加权MRI为20%(95% CI: 10, 35)(P < .001);sCT检测关节强直的敏感度为93%(95% CI: 78, 98),T1加权MRI为70%(95% CI: 47, 87)(P = .001)。
临床应用与阅片要点
- 在疑似骶髂关节炎的结构病变评估中,sCT对侵蚀、硬化和关节强直的检出准确率均高于常规T1加权MRI。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Evaluation of structural lesions in the sacroiliac (SI) joints can improve the accuracy for diagnosis of spondyloarthritis. However, structural lesions, such as erosions, are difficult to assess on routine T1-weighted MRI scans. Purpose To determine the diagnostic performance of MRI-based synthetic CT (sCT) in the depiction of erosions, sclerosis, and ankylosis of the SI joints compared with T1-weighted MRI, with CT as the reference standard. Materials and Methods A prospective study (clinical trial registration no. B670201837885) was performed from February 2019 to November 2019. Adults were referred from a tertiary hospital rheumatology outpatient clinic with clinical suspicion of inflammatory sacroiliitis. MRI and CT of the SI joints were performed on the same day. SCT images were generated from MRI scans using a commercially available deep learning-based image synthesis method. Two readers independently recorded if structural lesions (erosions, sclerosis, and ankylosis) were present on T1-weighted MRI, sCT, and CT scans in different reading sessions, with readers blinded to clinical information and other images. Diagnostic performance of sCT and T1-weighted MRI scans were analyzed using generalized estimating equation models, with consensus results of CT as the reference standard. Results Thirty participants were included (16 men, 14 women; mean age, 40 years ± 10 [standard deviation]). Diagnostic accuracy of sCT was higher than that of T1-weighted MRI for erosion (94% vs 86%, P = .003), sclerosis (97% vs 81%, P < .001), and ankylosis (92% vs 84%, P = .04). With sCT, specificity for erosion detection (96% [95% CI: 90, 98] vs 89% [95% CI: 81, 94], P = .01] and sensitivity for detection of sclerosis [94% [95% CI: 87, 97] vs 20% [95% CI: 10, 35], P < .001] and ankylosis (93% [95% CI: 78, 98] vs 70% [95% CI: 47, 87], P = .001) were improved. Conclusion With CT as the reference standard, synthetic CT of the sacroiliac joints has better diagnostic performance in the detection of structural lesions in individuals suspected of having sacroiliitis compared with routine T1-weighted MRI. © RSNA, 2020 Online supplemental material is available for this article. See also the editorial by Fritz in this issue.
引用
- Jans LBO, Chen M, Elewaut D, et al. MRI-based Synthetic CT in the Detection of Structural Lesions in Patients with Suspected Sacroiliitis: Comparison with MRI. Radiology. 2020. PMID: 33350891. doi:10.1148/radiol.2020201537
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。