文献速递妇产影像技术与安全MR

压缩感知与深度学习重建用于女性盆腔MRI去噪:在常规临床实践中提升图像质量与缩短检查时间的效用

Compressed sensing and deep learning reconstruction for women's pelvic MRI denoising: Utility for improving image quality and examination time in routine clinical practice.
PubMed 文献 · Eur J Radiol · 2020-11-21

连续30例女性盆腔疾病患者行T2加权成像,比较压缩感知(Compressed SPEEDER)与常规并行成像(SPEEDER)联合或不联合AiCE重建的检查时间、信噪比和图像质量。压缩感知联合AiCE可显著缩短检查时间并改善图像质量,但仅评估了T2加权序列。

要点

  • 压缩感知联合深度学习重建可显著缩短女性盆腔MRI的T2加权成像检查时间并改善图像质量。
  • 压缩感知联合AiCE与常规并行成像联合AiCE在信噪比和图像质量上无显著差异,但检查时间更短。
  • 压缩感知单独使用(不联合AiCE)的图像质量和信噪比显著低于其他重建方案。

核心结论

  • 在女性盆腔MRI的T2加权成像中,与常规并行成像(SPEEDER)相比,使用压缩感知(Compressed SPEEDER)联合AiCE可显著改善图像质量并缩短检查时间,但其他序列未进行测试。

研究设计与样本

  • 研究纳入30例连续女性盆腔疾病患者(平均年龄50岁),所有患者均行T2加权成像,分别采用压缩感知(Compressed SPEEDER)和常规并行成像(SPEEDER),并分别联合或不联合AiCE重建,共四组数据;记录检查时间并计算每例患者的信噪比,由两名阅片者共识采用5分制评估整体图像质量,采用Wilcoxon符号秩检验比较四组数据的平均检查时间、信噪比和整体图像质量。

关键影像表现/征象

  • 压缩感知联合或不联合AiCE的检查时间均显著短于常规并行成像联合或不联合AiCE(联合AiCE:p < 0.0001;不联合AiCE:p < 0.0001)。
  • 压缩感知联合AiCE及常规并行成像联合AiCE的信噪比均显著优于压缩感知不联合AiCE(压缩感知联合AiCE vs 压缩感知不联合AiCE:p = 0.01;常规并行成像联合AiCE vs 压缩感知不联合AiCE:p = 0.009)。
  • 压缩感知联合AiCE及常规并行成像联合或不联合AiCE的整体图像质量均显著高于压缩感知不联合AiCE(vs 压缩感知联合AiCE:p < 0.0001;vs 常规并行成像联合AiCE:p < 0.0001;vs 常规并行成像不联合AiCE:p = 0.0003)。

诊断标准与关键数值

  • 整体图像质量采用5分制评分,最终评分由两名阅片者共识确定。
  • 信噪比(SNR)对每例患者进行计算。

临床应用与阅片要点

  • 在女性盆腔MRI的T2加权成像中,采用压缩感知联合AiCE重建可在缩短检查时间的同时获得优于压缩感知不联合AiCE的图像质量。

局限性

  • 本研究仅对T2加权序列进行了测试,其他序列未进行测试。
  • 样本量为30例连续患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To demonstrate the utility of compressed sensing with parallel imaging (Compressed SPEEDER) and AiCE compared with that of conventional parallel imaging (SPEEDER) for shortening examination time and improving image quality of women's pelvic MRI.

METHOD: Thirty consecutive patients with women's pelvic diseases (mean age 50 years) underwent T2-weighted imaging using Compressed SPEEDER as well as conventional SPEEDER reconstructed with and without AiCE. The examination times were recorded, and signal-to-noise ratio (SNR) was calculated for every patient. Moreover, overall image quality was assessed using a 5-point scoring system, and final scores for all patients were determined by consensus of two readers. Mean examination time, SNR and overall image quality were compared among the four data sets by Wilcoxon signed-rank test.

RESULTS: Examination times for Compressed SPEEDER with and without AiCE were significantly shorter than those for conventional SPEEDER with and without AiCE (with AiCE: p < 0.0001, without AiCE: p < 0.0001). SNR of Compressed SPEEDER and of SPEEDER with AiCE was significantly superior to that of Compressed SPEEDER without AiCE (vs. Compressed SPEEDER, p = 0.01; vs. SPEEDER, p = 0.009). Overall image quality of Compressed SPEEDER with AiCE and of SPEEDER with and without AiCE was significantly higher than that of Compressed SPEEDER without AiCE (vs. Compressed SPEEDER with AiCE, p < 0.0001; vs. SPEEDER with AiCE, p < 0.0001; SPEEDER without AiCE, p = 0.0003).

CONCLUSION: Image quality and shorten examination time for T2-weighted imaging in women's pelvic MRI can be significantly improved by using Compressed SPEEDER with AiCE in comparison with conventional SPEEDER, although other sequences were not tested.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Ueda T, Ohno Y, Yamamoto K, et al. Compressed sensing and deep learning reconstruction for women's pelvic MRI denoising: Utility for improving image quality and examination time in routine clinical practice. Eur J Radiol. 2020. PMID: 33276249. doi:10.1016/j.ejrad.2020.109430

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。