辐射剂量降低对第三代双源CT肺动脉血管成像诊断性能的影响
研究纳入52例疑似肺栓塞行CT肺动脉造影的成人患者,用离线软件模拟25%、50%、75%原始剂量的低剂量图像,比较滤波反投影与迭代重建ADMIRE 3/5的图像质量和诊断准确性。以原始剂量CTPA为参考标准,患者水平迭代重建在各剂量下敏感度均为100%,ADMIRE 3/5在25%剂量(1.45 mSv)仍保持诊断准确性,FBP在25%剂量时患者水平敏感度降至93%。
要点
- 迭代重建ADMIRE 3/5在25%原始剂量(1.45 mSv)下患者水平敏感度100%、阴性预测值100%。
- FBP在50%和75%剂量下患者水平诊断准确率完美,25%剂量时敏感度降至93%、阴性预测值97%。
- 段水平ADMIRE 3/5在剂量降至25%时敏感度90.4%、特异度99.5%,为各方案中最低者见于FBP 25%剂量(敏感度84.6%、特异度98.9%)。
关键影像表现/征象
- 全部52次扫描的中位剂量长度乘积为291.1 ± 210.1 mGy·cm,有效剂量为5.8 ± 3.4 mSv。
- 52例中15例(28.8%)存在急性肺动脉栓塞。
- 总体主观图像质量在ADMIRE 5、75%原始剂量时最高,在FBP、25%原始剂量时最低(中位数[四分位距]:5[5] vs 3[2-3],p < 0.001)。
诊断标准与关键数值
- 患者水平诊断准确性:迭代重建数据集(ADMIRE 3和5)在各剂量下均为敏感度100%、阴性预测值100%。
- 患者水平诊断准确性:FBP在50%和75%原始剂量下诊断准确率完美,25%原始剂量时敏感度93%、阴性预测值97%。
- 段水平诊断准确性:ADMIRE 3和5在剂量降至25%时敏感度90.4%、特异度99.5%;FBP在25%剂量降低时敏感度84.6%、特异度98.9%。
- 肺栓塞检出的组内相关系数在所有剂量和重建方式下几乎完美(ICC:96.1-1.0,原文如此)。
- ADMIRE 3和5数据集在25%原始剂量(1.45 mSv)下、FBP在50%原始剂量(2.9 mSv)下可准确诊断肺栓塞。
临床应用与阅片要点
- 在第三代双源CT上采用迭代重建ADMIRE 3或5可将CTPA辐射剂量降至原始剂量的25%(1.45 mSv)而维持肺栓塞诊断准确性。
- 采用FBP重建时,剂量降至原始剂量的25%可使患者水平敏感度降至93%、阴性预测值降至97%,段水平敏感度降至84.6%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the effects of radiation dose reduction on diagnostic accuracy and image quality of pulmonary angiography CT (CTPA) in adults with suspected pulmonary embolism (PE).
MATERIAL & METHODS: 52 consecutive patients received CTPA for suspected PE. Realistic low-Dose CT simulations were generated using an offline software (ReconCT, Siemens Healthineers, Forchheim, Germany), as either filter back projections (FBP) or iterative reconstruction as ADMIRE (strength 3 or 5) with 25 %, 50 % and 75 % of the original dose. To assess image quality (overall image quality, noise, artifacts, and sharpness) and diagnostic confidence, a five-point scale was used. Patient-based and segment-based diagnostic accuracy was calculated for Low-dose computed tomography (LDCT)-reconstruction with original dose CTPA as a standard of reference. Furthermore, effective radiation doses were calculated using a commercially available dose management platform (Radimetrics, Bayer HealthCare, Leverkusen, Germany).
RESULTS: Among 52 patients, a total of 15 patients (28.8 %) had acute pulmonary artery embolism. The median dose-length product and effective dose for all 52 scans were 291.1 ± 210.1 mGy⋅cm and 5.8 ± 3.4 mSv. Overall subjective image quality was highest for ADMIRE 5 with 75 % and lowest for FBP with 25 % of the original dose (median [interquartile range]:5 [5] vs. 3 [2-3], p < 0.001. Patient-based diagnostic accuracy was perfect for all iteratively reconstructed data sets (ADMIRE 3 and 5) (sensitivity: 100 %, negative predictive value [NPV]: 100 %). LDCT data sets with FBP had perfect diagnostic accuracy at 50 % and 75 % of the original dose, which however decreased at 25 % of the original dose (sensitivity: 93 %; [NPV]: 97 %). Segment-based diagnostic accuracy was high for ADMIRE 3 and 5 down to 25 % dose reduction (sensitivity: 90.4 % specificity: 99.5 %) and lowest for FBP with 25 % dose reduction (sensitivity: 84.6 %, specificity: 98.9 %). Inter-class correlation regarding the detection of PE was almost perfect at all doses and recons (ICC: 96.1-1.0). Thus, accurate diagnosis for PE was possible for ADMIRE 3 and 5 datasets with 25 % of the original dose (1.45 mSv) and for FBP with 50 % of the original dose (2.9 mSv).
CONCLUSION: Our findings indicate that radiation dose reduction down to 25 % (1.45 mSv) of the original data via iterative reconstruction algorithms on a 3rd generation Dual Source CT (DSCT) scanner maintained the diagnostic accuracy and image quality for the assessment of PE in CTPA.
引用
- Winkelmann MT, Walter SS, Stock E, et al. Effects of radiation dose reduction on diagnostic performance of 3rd generation Dual Source CT pulmonary angiography. Eur J Radiol. 2020. PMID: 33254062. doi:10.1016/j.ejrad.2020.109426
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。