两点Dixon T2加权成像在甲状腺相关眼病中的价值:与常规脂肪抑制成像在脂肪抑制质量和分期效能上的比较
研究纳入37例甲状腺相关眼病(TAO)患者和15例健康对照,比较两点Dixon T2WI与常规脂肪抑制T2WI的图像质量、脂肪抑制均匀性及分期效能。Dixon T2WI在总体图像质量评分、脂肪抑制均匀性评分及眼外肌信号强度比(EOM-SIR)上均显著高于脂肪抑制T2WI(均p<0.05)。活动期TAO的Dixon-EOM-SIR、Fat-sat-EOM-SIR及Dixon-IF-WF均显著高于非活动期。Dixon-EOM-SIR≥3.32单独诊断活动期的敏感度最高(75.0%),联合Dixon-IF-WF≥0.09可提高分期效能和特异度(AUC 0.872;特异度97.1%)。作者建议Dixon T2WI用于临床评估TAO患者。
要点
- 两点Dixon T2WI的总体图像质量评分、脂肪抑制均匀性评分及EOM-SIR均显著高于常规脂肪抑制T2WI(均p<0.05)。
- 活动期TAO的Dixon-EOM-SIR、Fat-sat-EOM-SIR及Dixon-IF-WF均显著高于非活动期(p<0.001、p<0.001、p=0.001)。
- Dixon-EOM-SIR≥3.32单独诊断活动期敏感度最高(75.0%);联合Dixon-IF-WF≥0.09时AUC 0.872、特异度97.1%。
核心结论
- 与常规脂肪抑制技术相比,两点Dixon T2WI图像质量更好,对TAO的分期敏感度和特异度均有提高。
- 作者建议在临床实践中使用Dixon T2WI评估TAO患者。
研究设计与样本
- 研究纳入37例TAO患者和15例健康对照,均接受冠状位两点Dixon T2WI和脂肪抑制T2WI扫描。
- 比较两种序列的定性参数(总体图像质量、脂肪抑制均匀性)和定量参数(眼外肌信号强度比EOM-SIR),并在Dixon图像上测量眶内脂肪水分数(IF-WF)。
- 比较活动期与非活动期TAO的Dixon-EOM-SIR、Fat-sat-EOM-SIR和Dixon-IF-WF值,并评估其对活动期的诊断效能。
关键影像表现/征象
- 在TAO组和对照组中,两点Dixon T2WI的总体图像质量评分、脂肪抑制均匀性评分及EOM-SIR值均显著高于脂肪抑制T2WI(均p<0.05)。
- 活动期TAO的Dixon-EOM-SIR(p<0.001)、Fat-sat-EOM-SIR(p<0.001)和Dixon-IF-WF(p=0.001)均显著高于非活动期TAO。
诊断标准与关键数值
- ROC曲线分析显示,Dixon-EOM-SIR≥3.32单独诊断活动期的敏感度最高(75.0%)。
- 联合Dixon-EOM-SIR≥3.32和Dixon-IF-WF≥0.09可提高分期效能和特异度(AUC 0.872;特异度97.1%)。
临床应用与阅片要点
- 作者建议在临床实践中使用Dixon T2WI评估TAO患者。
- Dixon-EOM-SIR≥3.32单独应用时敏感度最高(75.0%),联合Dixon-IF-WF≥0.09可提高特异度(97.1%)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare the two-point Dixon T2 weighted imaging (T2WI) with conventional fat-sat T2WI in fat suppression (FS) quality and staging performance for patients with TAO.
METHODS: We enrolled 37 thyroid-associated ophthalmopathy (TAO) patients and 15 healthy controls who underwent both coronal two-point Dixon and fat-sat T2WI. Qualitative (overall imaging quality, FS uniformity) and quantitative [signal intensity ratio of extraocular muscle (EOM-SIR)] parameters were assessed between the two-point Dixon T2WI and fat-sat T2WI. Additionally, water fraction of intraorbital fat (IF-WF) was measured on Dixon image. Dixon-EOM-SIR, Fat-sat-EOM-SIR and Dixon-IF-WF values were compared between active and inactive TAO groups, and the diagnostic efficiency for the active phase were evaluated.
RESULTS: Two-point Dixon T2WI showed significantly higher overall image quality score, FS uniformity score as well as EOM-SIR value than fat-sat T2WI in both TAO and control groups (all p < 0.05). Active TAOs had significantly higher Dixon-EOM-SIR (p < 0.001), Fat-sat-EOM-SIR (p < 0.001) and Dixon-IF-WF (p = 0.001) than inactive TAOs. ROC curves analyses indicated that Dixon-EOM-SIR ≥3.32 alone demonstrated the highest staging sensitivity (75.0%). When integrating Dixon-EOM-SIR ≥3.32 and Dixon-IF-WF ≥0.09, improved staging efficiency and specificity could be achieved (area under the curve, 0.872; specificity, 97.1%).
CONCLUSION: Compared with conventional fat-sat technique, two-point Dixon T2WI offers better image quality, as well as improved staging sensitivity and specificity for TAO. Dixon T2WI is suggested to be used to evaluate the patients with TAO in clinical practice.
ADVANCES IN KNOWLEDGE: Two-point Dixon T2WI offers better image quality than fat-sat T2WI. Dixon-EOM-SIR alone demonstrated the highest staging sensitivity. Combining with Dixon-IF-WF showed improved staging efficiency and specificity. Dixon T2WI is suggested to be used to evaluate TAO patients in clinical practice.