超声自动ALARA方法:一项产科超声可行性研究
本研究在35名孕中期健康志愿者中评估基于实时图像质量反馈(滞后一相干性,LOC)的自适应超声方法。系统在0.15至1.20的16个机械指数(MI)值下扫描胎盘或胎儿腹部,将LOC-MI数据拟合S形曲线,以拟合曲线最大值的98%处对应的MI作为ALARA MI。结果显示ALARA MI值在0.35至1.03之间,与默认MI 0.8相比,80%的预试验采集建议更低的ALARA MI;LOC的拟合R²显著高于其他指标(P < .017)。结论支持基于LOC的自动ALARA方法在产科超声中的可行性。
要点
- 基于LOC的自适应方法可在产科超声中自动选择ALARA MI,其值在0.35至1.03之间。
- 与默认MI 0.8相比,80%的预试验采集建议更低的ALARA MI。
- LOC随MI变化的S形拟合R²显著高于其他图像质量指标(P < .017)。
诊断标准与关键数值
- ALARA MI定义为LOC随MI变化的S形拟合曲线达到其最大值98%时对应的MI值。
- 本研究中ALARA MI值在0.35至1.03之间,取决于声窗。
- 与默认MI 0.8相比,80%的预试验采集建议更低的ALARA MI。
- 对比度、对比噪声比、广义对比噪声比和LOC均随MI增加呈S形趋势;LOC的拟合R²显著高于其他指标(P < .017)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Ultrasound users are advised to observe the ALARA (as low as reasonably achievable) principle, but studies have shown that most do not monitor acoustic output metrics. We developed an adaptive ultrasound method that could suggest acoustic output levels based on real-time image quality feedback using lag-one coherence (LOC).
METHODS: Lag-one coherence as a function of the mechanical index (MI) was assessed in 35 healthy volunteers in their second trimester of pregnancy. While imaging the placenta or the fetal abdomen, the system swept through 16 MI values ranging from 0.15 to 1.20. The LOC-versus-MI data were fit with a sigmoid curve, and the ALARA MI was selected as the point at which the fit reached 98% of its maximum.
RESULTS: In this study, the ALARA MI values were between 0.35 and 1.03, depending on the acoustic window. Compared to a default MI of 0.8, the pilot acquisitions suggested a lower ALARA MI 80% of the time. The contrast, contrast-to-noise ratio, generalized contrast-to-noise ratio, and LOC all followed sigmoidal trends with an increasing MI. The R2 of the fit was statistically significantly greater for LOC than the other metrics (P < .017).
CONCLUSIONS: These results suggest that maximum image quality can be achieved with acoustic output levels lower than the US Food and Drug Administration limits in many cases, and an automated tool could be used in real time to find the ALARA MI for specific imaging conditions. Our results support the feasibility of an automated, LOC-based implementation of the ALARA principle for obstetric ultrasound.
引用
- Flint K, Bottenus N, Bradway D, et al. An Automated ALARA Method for Ultrasound: An Obstetric Ultrasound Feasibility Study. J Ultrasound Med. 2020. PMID: 33289152. doi:10.1002/jum.15570